Page 46
Electrical Installation
thermoIMAGER TIM
Fig. 20 Configuration process interface (PIF)
The standard process interface offers the following inputs and outputs:
Designation
Description
max. range
1/
status
AI
Analog input
0 - 10 V
2
DI
Digital input
24 V
AO
Analog output
0 - 10 V
Alarm output
0/ 10 V
The voltage levels for the DI are: LOW = 0 ... 0.6 V / HIGH = 2 ... 24 V.
6.3
Industrial Process Interface (Optional)
For use in industrial environment an industrial process interface with 500 VAC
RMS
isolation voltage between
TIM and process is available (connection box with IP 65, 5 m, 10 m or 20 m standard or high temp cable for
camera connection, terminal for process integration)
(Industrial Process Interface).
6.4
USB Cable Extensions
The maximum USB cable length is 20 m. For greater distances between TIM and computer or for stand-alone
solutions you should use the optional TIM NetBox or the USB-Server Industry Isochron
1) Depending on supply voltage; for 0 - 10 V on the AO the PIF has to be powered with min. 12 V.
2) The AI is designed for max. 24 V, the voltage level above 10 V is not interpreted.