Page 3

General

IMS5x00

General

Symbols used

The following symbols are used in this document:

CAUTION

Indicates a hazardous situation which, if not avoided, may result in minor or moderate injury.

NOTICE

Indicates a situation that may result in property damage if not avoided.

Indicates a user action.

i

Indicates a tip for users.

Measure

Indicates hardware or a software button/menu.

Warnings

CAUTION

Connect the power supply and the display/output device according to the safety regulations for 
electrical equipment.

 

> Risk of injury, damage to or destruction of the controller

NOTICE

 The supply voltage must not exceed the specified limits.

 

> Damage to or destruction of the controller
Avoid shocks and impacts to the sensor and the controller.

 

> Damage to or destruction of the components
Never kink optical fibers or bend them in thight radii.

 

> Damage to or destruction of the optical fibers; failure of measurement device
Protect the ends of the optical fibers against contamination

 

> Failure of the measuring device
Protect the cable against damage.

 

> Failure of the measuring device

Содержание interferoMETER IMS5400-DS

Страница 1: ...IMS5400 DS IMS5600 DS IMS5400 TH Quick Manual interferoMETER ...

Страница 2: ...ions 6 Sensor Cable 7 Mounting Mounting Adapter 7 Pluggable Screw Terminals 8 LEDs Controller 9 Button Multifunction 10 Initial Operation 11 Select a Sensor 11 Start of Measuring Range Working Distance 11 Positioning the Target Distance Measurement 12 Positioning the Target Thickness Measurement 13 Material Selection 13 Measurement Configuration 14 FFT Signal Check 15 Signal Quality 15 Distance an...

Страница 3: ...ay output device according to the safety regulations for electrical equipment Risk of injury damage to or destruction of the controller NOTICE The supply voltage must not exceed the specified limits Damage to or destruction of the controller Avoid shocks and impacts to the sensor and the controller Damage to or destruction of the components Never kink optical fibers or bend them in thight radii Da...

Страница 4: ...and other material goods are damaged in the event of malfunction or total failure of the controller Take additional precautions for safety and damage prevention in case of safety related applications Proper Environment Protection class sensor IP40 with connected sensor cable only controller IP40 Lenses are excluded from protection class Contamination of the lenses causes impairment or failure of t...

Страница 5: ...ssible radiation is harmless under predictable con ditions For class 1 laser devices impairment of color vision and disturbances e g from a glare effect cannot be excluded An LED signalizes by illumination that laser radiation emits from the optical opening of the light source Pilot on Class 1 Laser Product IEC 60825 1 2014 P 0 01 mW 635 nm COMPLIES WITH 21 CFR 1040 10 AND 1040 11 EXCEPT FOR CONFO...

Страница 6: ...mm 1 Measuring ranges for distance and thickness measure ments The possible resolution here is in the nanometer range For a quick start we recommend to use presets defined for different target surfaces and applications see operat ing instructions Chap 6 6 1 Measuring range with n 1 5 for air gap measure ment between two glass plates n 1 the measuring range is 0 05 2 1 mm The measurement object mus...

Страница 7: ...s any mechanical stress of the optical fiber bending crushing pulling twisting knotting etc tight curvature of the optical fiber because the glass fiber is damaged in the process and this causes permanent damage Never bend the cable more tightly than the permissible bending radius R Fixed R 30 mm or more Flexible R 40 mm or more Mounting Mounting Adapter The IMP DS19 TH45 series sensors use an opt...

Страница 8: ...al I O GND Shield TrigIn GND Error 1 GND Shield Error 2 GND TrigIn Trigger input TTL or HTL level TTL Low 0 8 V High 2 V HTL Low 3 V High 8 V Error 1 2 Switch outputs NPN PNP or Push Pull I max 100 mA U H max 30 V GND Ground potentials All GND conductors are interconnected with one another and to operating voltage ground 24 VDC Operating voltage 15 I max 1 A 24 VDC GND Shield Power GND Operating v...

Страница 9: ...easuring range Green Target within measuring range SLED Red SLED of Yellow SLED warms up Green SLED ready for operation Yellow flashing SLED current outside the optimal value range 1 Pilot Red Pilot laser off Green Pilot laser on Green Pilot laser is alternately turned on and off if no target object or outside the measuring range 1 When measuring outside the optimum current value of the SLED the c...

Страница 10: ...s the master measurement of the selected signals Pilot laser Turns on off the pilot laser SLED Turns the light source on off for the sensor Inactive Key has no function There are two defined time intervals for pressing the button each of these can be assigned a function All time inter vals are indicated by the LEDs flashing lighting up Pilot laser SLED Master value Function 1 2 sec 0 5 sec 10 sec ...

Страница 11: ... in the web browser now You can check the IP address of the controller that are connected to a PC network with the sensorTOOL exe program This program is available online at https www micro epsilon de download software sensorTOOL exe Start the program sensorTOOL and click the button Click the Open WebPage button to connect the controller to your default browser Select a Sensor Change to the Settin...

Страница 12: ...d off alternately if no target or outside the measuring range You can also position the sensor using the FFT sig nal 1 The interferometric measuring principle provides measurement values in front of and behind the actual measuring range An incorrect measuring range dis tance can be recognized by the running direction of the peak in the FFT signal Inverse direction of FFT signal if the target is ou...

Страница 13: ...ds reduces the materials database For a new material the group refractive index is required 1 The maximum thickness for a air gap is 2 1 mm The thickness for glass n 1 5 is 35 µm as a minimum and 1 4 mm as a maximum Positioning the Target Thickness Measurement The red light pilot laser supports you in aligning the sensor to the target during commissioning Turn on or off the pilot laser in the menu...

Страница 14: ...d preset the web interface displays a dialog which enables the user to define a setup name to avoid accidental overwriting This prevents presets from being over written by accident Distance measurement 1 e g on ceramics non transparent plastics No averaging Individual material selection is possible in the Set tings Data record ing Material selec tion menu 1 The distance presets are only possible w...

Страница 15: ...ine peak marking The diagram starts automatically when the website is called Signal Quality A good measurement result can be achieved with sufficient FFT signal intensity Reducing the measuring rate enables longer exposure of the CCD array therefore leading to high measurement quality Go to the menu Home Signal quality and adapt the measurement dynamics to the requirements Check the result in the ...

Страница 16: ...lly to the target object Then move the sensor or the target closer until you more or less reach the start of measuring range for your sensor Once the object is within the sensor s measuring range the Range LED green or yellow on the front of the controller will light up Or observe the FFT signal 2 1 3 4 5 6 2 8 7 Measurement web page distance measurement ...

Страница 17: ...are gray Click on the check mark to add them The changes take effect when saving the settings Use the eye symbols to show and hide the single signals The calcula tion continues in the background 01PEAK01 Chronological sequence of dis placement signal 4 Auto automatic scaling or Manual manual setting allow for scaling the measure ment axis Y axis of the graphic 5 The current values for distance exp...

Страница 18: ...te Switching limit value output Analog output outputs either voltage or current values Selecting the required interfaces for data output Switch to the Settings Outputs Output interface menu and select the desired output channels Ethernet The controller transmits TCP IP or UDP IP packages with an Ethernet transfer rate of 10 Mbit s or 100 Mbit s The transfer rate is selected automatically depending...

Страница 19: ...must be notified immediately The liability for material defects is 12 months from deliv ery Within this period defective parts except for wear ing parts will be repaired or replaced free of charge if the device is returned to MICRO EPSILON with shipping costs prepaid Any damage that is caused by improper handling the use of force or by repairs or modifica tions by third parties is not covered by t...