Page 57
Operation
induSENSOR DTD / MSC7xxx
5.4.2
Zero-point Adjustment
D1:
Channel
D2:
E1
D3:
E2
D4:
Value
G
O
LED off
Set the output signal
V
0
.
6 VDC or 12 mA is preset.
ENTER
G
R
G
O
The LED flashes
and color changes
depending on the
output signal
(green = too low
red = too high).
O
Move the measuring object to position
X
0
until the
output has reached
V
0
.
Lights orange when
the measuring object
is in the electrical
center of the sensor.
ENTER
G
O
The LED color
changes depending
on the position
of the measuring
object.
O
Move the measuring object to position
X
2
1
and
change the output signal
V
2
with
Flashes orange when
the measuring object
is in the electrical
center of the sensor.
Fig. 59 Menu structure for the MSC 7401 / 7802 / 7602 controllers, adjustment mode: Zero-point search
1) Position
X
2
must
be > 10 % of the
measuring range
away from
X
1
.