Page 100
Warranty
optoNCDT 1320
A 4
Control Menu
A 4.1
Tab Home
Measurement
task
Presets
Standard
Suitable for materials made of ceramics, metal or filled plastics
Multi surface
1
Suitable for e.g. PCB or hybrid materials
Light penetration
1
Suitable for plastics (POM, Teflon), materials with strong pen-
etration of the laser
Setup
Setup 1
A setup contains user-specific measurement settings. Unlike the
presets it can be changed anytime.
Signal quality
Static / balanced / dynamic
The signal quality affects averaging of measurement values.
A 4.2
Tab Settings
A 4.2.1
Inputs
Laser on/off
On / Off
Laser on/off is only effective when pin 8 is connected to GND.
Multifunctional input
Zeroing (Mastering) High / Low Sets the function of the switching input. The trigger affects
recording and output of a measurement value. Zeroing/Master-
ing sets the current measurement value to the entered master
value. Teaching scales the analog output. HTL is defined as
active input level.
Trigger In
High / Low
Teaching
Inactive
Key function
Zeroing (Mastering)
Sets the function of the sensor key. Inactive means keylock.
Teaching
Inactive
1) Available for the sensor models ILD1320-10/25/50.