eddyNCDT 306x / 307x
Measurement Setup, Operating Multiple Sensors
Sensors of the eddyNCDT 306x, 307x series cannot be synchronized. Observe the following installation information
regarding the minimum distance between two sensors:
- 3x sensor diameter distance between two unshielded sensors with equal carrier frequency (e. g. low frequency)
- 1.5x sensor diameter distance between two shielded sensors with equal carrier frequency (e. g. low frequency)
- two nearby mounted sensors only as low frequency and high frequency models
low frequency
high frequency
Target
low frequency
ES-Ux
low frequency
Target
min 3 x
sensor dia
.
Target
low frequency
high frequency
low frequency
high frequency
low frequency
low frequency
Target
low frequency
ES-Sx
low frequency
Target
min 1.5 x
sensor dia.
high frequency
high frequency
Target
Sensor, ES-Ux
Sensor, ES-Sx
No synchronization required
Not possible