![MICRO-EPSILON confocalDT 2421 Скачать руководство пользователя страница 7](http://html.mh-extra.com/html/micro-epsilon/confocaldt-2421/confocaldt-2421_quick-manual_1785312007.webp)
Page 7
System Design, Connection Options
confocalDT 2421 / 2422
Screw Terminals
Pin
Description
Notes
U/I out
GN
D
Shield
Analog Out
U/I out
Voltage output
0 ... 5 V; 0 ... 10 V;
R
i
approx. 50 Ohm; 5.5 V / 10.9 V in the event of an
error, outside of measuring range
Current output
4 ... 20 mA; R
L
≤ 500 Ohm; 23.7 mA in the event of
an error, outside of measuring range
GND
Ground analog output Electrically connected to supply
+Sync/Trig
-Sync/Trig
Input/output synchroni-
zation, input triggering
RS422 level (EIA422)
+
Sync/T
rig
-
Sync/T
rig
Digital I/O
GND
Shield
TrigIn
GND
Er
ror 1
GND
Shield
Er
ror 2
GN
D
TrigIn
Input triggering
TTL or HTL level
TTL: Low ≤ 0.8 V, High ≥ 2 V
HTL: Low ≤ 3 V, High ≥ 8 V
Error 1 / 2
Error outputs
NPN, PNP or Push-Pull
I
max
= 100 mA, U
H max
= 30 V
GND
Reference ground
All GND conductors are interconnected with one
another and to operating voltage ground.
24 VDC
Operating voltage
± 15%, I
max
< 1 A
24 VDC
GN
D
Shield
Power
GND
Operating voltage
ground
GND is electrically connected to GND of switching
outputs, synchronization, analog and encoder input
Shield
Shields for relevant output/input, plug housing
The plug-in screw terminals are designed for an optical fiber cross-section of 0.14 mm² to 1.5 mm².