Page 40
Functional Principle, Technical Data
capaNCDT 6100
2.3
Technical Data
Controller type
DT6100
Resolution static
0.01 % FSO
Resolution dynamic
0.015 % d.M. (2 kHz)
Limit frequency
2 kHz
Limit frequency adjustable
10 Hz / 2 kHz
Linearity
±0.3 % d.M. (all sensors interchangeable without calibration)
Option LC: ±0.1 % FSO (tuned to one sensor)
Max. sensitivity deviation
±0.1 % FSO
Long term stability
≤ 0.05 % FSO / Month
Synchronous operation
yes
Insulator measurement
no
Temperature stability
±0.03 % FSO / °C
Operating temperature
Controller: +10 ... +60 °C (+50 to +140 °F)
Sensor: -50 ... +150 °C (-58 to +302 °F)
Storage temperature
-10 … +75 °C (+14 ... +167 °F)
Supply
24 VDC / 85 mA (9 ... 36 VDC)
optional ±15 VDC / 85 mA (9 ... 36 VDC)
Output
0 … 10 V (resistance min. 1.2 kΩ / capacitance max. 1 nF)
optional: 4 … 20 mA (load max. 400 Ω)
Sensors
all sensors außer CS005
Electromagnetic compatibility (EMC)
DIN EN 61326-1: 2006-10 and DIN 61326-2-3: 2007-05
FSO = Full Scale Output