Page 31
LED Functions
thermoMETER CSmicro
6.2
Self Diagnostic
With this function the current status of the sensor will be indicated by different flash modes of the LED.
If activated, the LED will show one out of five possible states of the sensor:
Status
LED mode
Normal
intermittent off
- - - -
Sensor overheated
fast flash
-------------------
Out of measuring range
double flash
-- -- -- -- -- -- --
Not stable
intermittent on
Alarm fault
always on
i
At a power supply (Vcc) ≥ 12 V it takes about 5 minutes until the sensor works in a stable mode.
Therefore, after switching on the unit, the LED will show a not stable state for up to 5 minutes.
Sensor overheated:
The internal temperature probes have detected an invalid high internal
temperature of the thermoMETER CSmicro.
Out of measuring range:
The object temperature is out of measuring range.
Not stable:
The internal temperature probes have detected an unequally internal tem-
perature of the thermoMETER CSmicro.
Alarm fault:
Current through the switching transistor of the open-collector output is to
high.
Содержание 2W-SF15H
Страница 6: ...thermoMETER CSmicro ...
Страница 21: ...Page 21 Optical Charts thermoMETER CSmicro SF15 2W 2W SF15H HS with CF lens 0 8 mm 10 mm 2W SF22H D S 22 1 ...
Страница 22: ...Page 22 Optical Charts thermoMETER CSmicro 2W SF22H with CF lens 0 6 mm 10 mm SF02 D S 2 1 ...
Страница 23: ...Page 23 Optical Charts thermoMETER CSmicro SF02 with CF lens 2 5 mm 23 mm M 2SF40 SF D S 40 1 ...
Страница 24: ...Page 24 Optical Charts thermoMETER CSmicro M 2SF40 CF D S 40 1 D S Far field 12 1 M SF75 SF D S 75 1 ...
Страница 27: ...Page 27 Optical Charts thermoMETER CSmicro M 3H SF D S 33 1 M 3H CF D S 33 1 D S Far field 11 1 ...
Страница 71: ......