20
8. Remote Flowhead
The Micro flowhead can be set up to work remotely from the device. This
may be useful if the display needs to be monitored while the subject uses
the device.
1. Slide the flowhead away from the device from left to right. You
may need to push firmly to get it to move.
2. From the remote flowhead adapter kit, attach the device cap in
the space that was occupied by the flowhead.
3. Attach the remote flowhead adapter to the flowhead. This is done
by sliding the flow head into the grooves in the remote flowhead
adapter. Ensure this is fully pushed in.
4. Attach the remote flowhead to the port on the Micro Spirometer
cap using the flowhead connection tube.
5. It is recommended that an accuracy check is carried out before
the flowhead is used remotely or refitted to verify correct
operation and accuracy.
9. Fault Finding Guide
Problem Fault
Symptoms:
•
Calibration verification variations > ± 3%.
•
False readings suspected.
Possible
Solutions: (In
probable order)
•
The % error indicates the severity of the issue.
Over 3%, repeat the calibration verification. Over
25%, the user should contact technical support.
•
Was the correct syringe volume selected?
•
Flowhead pressure tapping holes blocked.
Contact support.
•
Flowhead Fleisch element assembly sealing
‘O’
rings damaged. Contact support.
•
Flowhead Fleisch element assembly blocked.
Contact support.
•
Cold syringe. Ensure syringe is in its test
environment for at least an hour before use.
•
EMI (Electromagnetic Interference) that exceeds
EN60601-1-2 EMC standard. Turn off offending
device or move the Micro Spirometer to another
location and repeat.
•
Internal tubing from pressure ports on device is
blocked. Contact support.
Problem Fault
Symptoms:
•
Cannot print to external printer
Possible
Solutions: (In
•
Check device is connected to the computer and
Device Studio is installed on the computer.