18 

 

Battery Depleted 

– an orange depleted battery icon. 

Continued use is allowed, however prepare to change 
the batteries, or connect via USB cable to a computer 
to continue testing. 

 

Battery Discharged 

– the Battery Discharged icon 

displays on full screen on power up and a red empty 
battery icon displays on the status bar. Batteries 
should be changed immediately or the device 
connected to a computer via USB cable to continue 
testing. 

 

When the USB power is connected, this icon 

 displays instead of the 

battery icon. 

6.2 Power Save Mode 

To improve battery life, when powered by batteries, the Micro Spirometer 
will auto power down if left unused for 5 minutes. 

When running of the USB, the device screen will go blank if left unused 
for 5 minutes. The device will not auto power down if powered by USB. 

Press the screen or on/off button will bring the device out of power save 
mode. 

7. Cleaning & Hygiene 

7.1 Preventing Cross-Contamination of Subjects 

A spirometer is not designed or supplied as a 

‘sterile’ device. Micro Direct 

intends that a new SpiroSafe filter be used for every subject to prevent 
cross contamination. Using a new SpiroSafe provides a significant level 
of protection for the subject, the device and the user against cross 
contamination during spirometry maneuvers. 

The interior of the Micro flowhead does not require decontamination 
where a new SpiroSafe filter is used for each subject. The outside 
surfaces of the device and the flowhead tube may be cleaned with a 70% 
isopropyl alcohol impregnated cloth to remove any visible soiling and for 
low level disinfection. 

Where the user suspects the flowhead has become contaminated or 
where local risk assessment identifies the need for higher level of 
decontamination, then it should be cleaned as per the instructions on 
‘Cleaning and Hygiene’ on the Micro Direct website (www.mdspiro.com). 

Table of Cleaning/Disinfection Methods 

 

 

 

Содержание MD SPIRO Micro MD6300

Страница 1: ...Micro Model MD6300 Instruction for Use Current Edition Issue 2 29 Jun 2022 Cat No 09360 Micro Direct Inc 803 Webster Street Lewiston ME 04240 Toll Free 800 588 3381 Email support mdspiro com...

Страница 2: ...y disease Frequent colds Assessment of congestive heart failure CPT Codes for Spirometry 94010 Spirometry Complete Includes graphic record total and timed vital capacity expiratory flow rate measureme...

Страница 3: ...5 3 Reporting 12 5 4 Calibration Verification 13 5 5 Configuration Options 15 6 Power Management 17 6 1 Batteries 17 6 2 Power Save Mode 18 7 Cleaning Hygiene 18 7 1 Preventing Cross Contamination of...

Страница 4: ...SpiroSafe filter provides a significant level of protection of the subject the device and the user against cross contamination during spirometry maneuvers A SpiroSafe filter is for single use only 4...

Страница 5: ...e instrument is measuring accurately The Micro Spirometer should never be outside accuracy limits A calibration verification should be completed after cleaning or disassembling the spirometer after ad...

Страница 6: ...quipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation If such use is necessary this equipment and the other equipment should be observed...

Страница 7: ...Computer Shown for illustration purposes only and not supplied 1 USB Flash Drive containing Device Studio Software 2 Power On Off Button 3 Micro USB Port 4 Plinth 5 Flowhead 6 Flowhead Cone 7 LCD Touc...

Страница 8: ...ice Studio the Micro Spirometer must be connected to the computer using the USB cable via ports marked with the symbol If the device has just been unpacked or transported ensure it is left sitting ful...

Страница 9: ...whead Cone 3 Operating Instructions On first time use the Micro opens on the Setup screen for Date and Time Enter the current date and time Press the forward icon to save Continue to the temperature e...

Страница 10: ...oup are on the second screen and may be enabled in the Configuration Menu Subject Options 3 To enter the information for Date of Birth Height and Weight touch the blank field on the screen to open the...

Страница 11: ...ubject should Sit upright and maintain this posture throughout the test Fit the noseclip and relax Place SpiroSafe Filter in mouth and close lips around the mouthpiece Seal lips around the mouthpiece...

Страница 12: ...eft right arrows to select which test results to view The tests are shown in order of rank best is ranked number 1 then 2 3 etc 9 To delete the current blow Select the Delete option from the side menu...

Страница 13: ...to return to normal mode These are not enabled during test 10 The results summary at the top of the screen shows the FVC and FEV1 of the last blow 11 The number of tests for VC or the number of usabl...

Страница 14: ...Note A session ends and is saved when one of the following occur the device is turned off a new subject is created or the device is connected to Device Studio 5 2 3 Post Bronchodilator Post Test A Po...

Страница 15: ...and in the Main Menu 3 When connected the remote mode icon will display on the Micro Spirometer 4 Device Studio will search for assessments on the device The user will be presented with the download...

Страница 16: ...e a calibration verification 1 Select the Configuration icon on the Main Menu screen 2 Select the calibration verification icon 3 Enter the syringe volume serial number and ambient temperature using t...

Страница 17: ...or an issue with the flowhead device contact technical support Outside 3 6 purple syringe and shows 3 6 on the top of the screen The higher the the more severe the issue Over 3 repeat calibration ver...

Страница 18: ...if the flowhead is dropped or replaced and if EMI Electromagnetic Interference is suspected or possible 5 5 Configuration Options To access the Configuration menu press the icon on the Main Menu scre...

Страница 19: ...VC of FVC FEV 5 Forced Expiratory Volume after 0 5 seconds L PIF L s Peak Inspiratory Flow L sec FMFT Forced Mid Expiratory Flow Time sec FET Forced Expiratory Time sec FEV 5 FVC Ratio FEV 0 5 of FVC...

Страница 20: ...etails on performing a Calibration Verification 4 About Contains information about the software which should be used if making inquiries to the manufacturer or a service agent This information include...

Страница 21: ...he device out of power save mode 7 Cleaning Hygiene 7 1 Preventing Cross Contamination of Subjects A spirometer is not designed or supplied as a sterile device Micro Direct intends that a new SpiroSaf...

Страница 22: ...utine basis Examine cone and flow conditioning mesh filter for damage or contamination If damaged or blocked it should be replaced with a new part Examine the Fleisch element and replace if damaged If...

Страница 23: ...accuracy 9 Fault Finding Guide Problem Fault Symptoms Calibration verification variations 3 False readings suspected Possible Solutions In probable order The error indicates the severity of the issue...

Страница 24: ...g is connected correctly Electronics failure Contact support 9 1 Software Check Information about the device may be obtained from the About option in the Configuration Menu This information may be use...

Страница 25: ...ro com support mdspiro com Website www mdspiro com 9 Consumables and Accessories Cat No Description 3385 SpiroSafe Viral Bacterial Filters 100 3304 Nose Clips 20 3325 3 Liter Calibration Syringe 69131...

Страница 26: ...rating instructions Manufacturer Date of Manufacture Date format yyyy mm dd USB connector On Off The device must be taken to separate collection at the product end of life Do not dispose of these prod...

Страница 27: ...reports The Fleisch flowhead is used for testing and is integral to the device 13 Technical Specifications Product Micro Spirometer Model MD6300 Flow Detection Principal Fleisch Type Pneumotachograph...

Страница 28: ...ance Flow measurement output Essential Performance Test Limits Flow accuracy 10 or 10 L min with 3 or 3 L min allowed for test apparatus ISO 23747 Service Life The recommended service life of the devi...

Страница 29: ...Model MD6300 Micro Spirometer to the Medical Devices Directive of the European Community The Model MD6300 Micro Spirometer is intended for use in a variety of healthcare environments e g primary care...

Страница 30: ...ly for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The Model MD6300 micro is...

Страница 31: ...terference is evident or possible Loss or degraded performance due to EMI that exceeds the test levels in immunity test tables above will result in a failed calibration verification on the Micro Spiro...

Страница 32: ...s IIa per Annex IX of the MDD also meets the provision of the Essential Requirements Annex I via compliance with Annex II of the Medical Devices Directive as per Article 11 section 3a excluding point...

Страница 33: ...ny s literature and user manuals The Company undertakes to rectify at no expense to the customer any software failure notified within the period stated above provided the failure can be recreated and...

Страница 34: ...0 J30 5 Asthma Mild Intermittent J45 20 J45 22 Asthma Mild Persistent J45 30 J45 32 Asthma Moderate Persistent J45 40 J45 42 Asthma Severe Persistent J45 50 J45 52 Asthma Unspecified J45 901 J45 909 C...

Страница 35: ...espiratory conditions due to unspecified external agent J70 9 Sarcoidosis of the Lung D86 0 Sarcoidosis of the Lung with sarcoidosis of the lymph nodes D86 2 Bronchiolitis Acute J21 0 J21 9 Bronchitis...

Страница 36: ...rperium O99 330 O99 335 Wheezing R06 2 Use additional code after the primary diagnosis to identify any tobacco use dependence or exposure to tobacco smoke Please Note Information in this manual is sub...

Отзывы: