REPLACING PARTS
4. Thoroughly clean the surfaces of the housing body to prevent contamination
entering the inner assemblies.
5. Align and install the new zirconia sensing module (6) to the sensor head (1).
Gently push to secure the module in place: refer to Figure 14.
NOTE:
The connector is keyed to aid alignment.
!
CAUTION: Do NOT force the module otherwise you may damage
the terminals or other components within the housing.
Figure 14
Insert Sensor block
6. Replace the sensor head as described in Section 9.2.
7. Reapply the power.
8. Calibrate the new sensing module: refer to section 6.2.
9.4
Removing / Replacing the Front Panel
Refer to Figure 2.
1. If fitted, carefully remove the two screw covers (2) from the front panel (1).
2. Remove the four screws securing the front panel (1) to the housing
3. Carefully ease the front panel (1) from the housing.