XZR250 User’s Manual
8
97528
Issue 5, May 2019
INSTALLATION
2.4
Probe Installation
I
!
CAUTION: Do not mount the probe at an angle that would allow
liquid to run down towards the sensor head, as this may cause
blockage or damage.
NOTE:
A suitable 2” ANSI flange mounting adaptor is required. If the application temperatures
will exceed 200°C, use an anti-seize compound on any mounting threads to ease future
removal of the probe.
1. Cut a 50mm (2” nom. pipe) diameter hole in the flue wall (or other mounting position).
2. Either:
•
Weld the mounting adaptor into place, concentric with the hole in the flue.
•
Drill and bolt the mounting adaptor to the flue.
3. Before bolting in place, fit a suitable gasket between the adaptor and the flue wall.
A gasket is supplied with each analyzer. (Order code:
XZR250A1-GASKET01)
4. Carefully insert the probe into the hole in the flue (or other mounting position).
5. Ensure the probe is in the correct orientation using the flow direction indicator on
the probe's flange.
6. Secure the probe into place using suitable bolts and washers.
Figure 8
Probe Installation
!
CAUTION: Always install a suitable gasket between the mating
flanges to prevent gas leakage. The wrong type of gasket may
lead to accidental poisoning of the zirconia sensor: refer to 5.1.3