XTP501 & XTC501 User’s Manual
Michell Instruments
35
INSTALLATION
4.5
Electrical Installation
4.5.1 Power Supply and Input/Output Signal
The 501 requires 24 V DC power input at a maximum start-up current of 1.5 A.
NOTE : Loose connectors are supplied with the analyser for Power, Inputs,
Outputs and Alarms.
Figure 32
Connections
4.5.2 Power Supply
POWER FEED
Power
Pin Number
NC
1
24 V ±4 V
2
NC
3
0 V
4
NOTE: The above view of the electrical
connections is shown from the mating side.
Here are the connections as viewed from the back of the
connector and this should be used for wiring purposes.
Содержание XTC501
Страница 1: ...XTP501 Oxygen Analyzer XTC501 Binary Gas Analyzer User s Manual 99976 Issue 1 1 May 2020...
Страница 18: ...XTP501 XTC501 User s Manual 10 99976 Issue 1 1 May 2020 OPERATION 2 4 1 Menu Map Figure 9 Menu map...
Страница 47: ...XTP501 XTC501 User s Manual Michell Instruments 39 APPENDIX A Appendix A Technical Specifications...
Страница 56: ...http www michell com...