![Michell Instruments Easidew PRO I.S. Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/michell-instruments/easidew-pro-i-s/easidew-pro-i-s_user-manual_1784702030.webp)
Easidew PRO I.S. User’s Manual
22
97130 Issue 15, May 2018
APPENDIX C
C.5 Special
Conditions
1. The wiring connections to the free socket must be made via crimped
connectors in such a way that all the strands of the wire used are held
securely by the crimp.
2.
The plastic plug and socket create a potential for electrostatic discharge
so must not be rubbed with a dry cloth or cleaned with solvents.
3.
The Easidew PRO I.S Dew-Point Transmitter does not withstand the 500
V AC insulation test to frame. This must be taken into account when
installing the equipment.
C.6
Maintenance and Installation
The Easidew PRO I.S. must only be installed by suitably quali
fi
ed personnel and in
accordance with the instructions provided and the terms of the applicable product
certi
fi
cates.
Maintenance and servicing of the product must only be carried out by suitably trained
personnel or returned to an approved Michell Instruments’ Service Center.
Содержание Easidew PRO I.S.
Страница 20: ...Easidew PRO I S User s Manual 12 97130 Issue 15 May 2018 APPENDIX A Appendix A Technical Specifications...
Страница 24: ...Easidew PRO I S User s Manual 16 97130 Issue 15 May 2018 APPENDIX B Appendix B System Drawings...
Страница 26: ...Easidew PRO I S User s Manual 18 97130 Issue 15 May 2018 APPENDIX B B 2 FM Approved System Drawing...
Страница 28: ...Easidew PRO I S User s Manual 20 97130 Issue 15 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 35: ......
Страница 36: ...http www michell com...