OPERATION
Minimize Dead Space in Sample Lines:
Dead space causes moisture entrapment points, increased system response times and
measurement errors, as a result of the trapped moisture being released into the passing
sample gas and causing an increase in partial vapor pressure.
Deadspace
i
Figure 9
Indication of Dead Space
Remove Any Particulate Matter or Oil from the Gas Sample:
Particulate matter at high velocity can damage the sensing element and similarly, at
low velocity, they may ‘blind’ the sensing element and reduce its response speed. If
particulate, such as degraded desiccant, pipe scale or rust is present in the sample gas,
use an in-line
fi
lter, as a minimum level of protection. For more demanding applications
Michell Instruments offers a range of sampling systems (for more information contact
www.michell.com).
Use High Quality Sample Tube and Fittings:
Michell Instruments recommends that, wherever possible, stainless steel tubing and
fi
ttings should be used. This is particularly important at low dew points since other
materials have hygroscopic characteristics and adsorb moisture on the tube walls,
slowing down response and, in extreme circumstances, giving false readings. For
temporary applications, or where stainless steel tubing is not practical, use high quality
thick walled PTFE tubing.
Position Transmitter away from Heat Source:
It is recommended, as good instrumentation practice, that the transmitter is placed
away from any heat source to avoid adsorption/desorption.
Содержание Easidew PRO I.S.
Страница 20: ...Easidew PRO I S User s Manual 12 97130 Issue 15 May 2018 APPENDIX A Appendix A Technical Specifications...
Страница 24: ...Easidew PRO I S User s Manual 16 97130 Issue 15 May 2018 APPENDIX B Appendix B System Drawings...
Страница 26: ...Easidew PRO I S User s Manual 18 97130 Issue 15 May 2018 APPENDIX B B 2 FM Approved System Drawing...
Страница 28: ...Easidew PRO I S User s Manual 20 97130 Issue 15 May 2018 APPENDIX C Appendix C Hazardous Area Certification...
Страница 35: ......
Страница 36: ...http www michell com...