34
CAUTION
Take extreme care that the liquid does not soak into the data ports. Do not
use any type of spray.
• Dispose of the used wipes.
• Wash your hands thoroughly with soap and water. The meter must be cleaned
and disinfected a minimum of once per week.
CAUTION
If the meter is being operated by a second person who is providing
testing assistance to the user, the meter and lancing device should be
decontaminated prior to use by the second person.
Reusable lancing device
1. Wash your hands before performing cleaning and disinfecting procedures.
2. Remove the wipe from the container.
3. Make sure the used lancet is removed prior to cleaning.
4. Wipe the outside of the reusable lancing device carefully (3 passes horizontal
and 3 passes vertical).
* To clean the reusable lancing device, wipe it with a CaviWipes™ towelette.
* To disinfect the reusable lancing device, wipe it with an other CaviWipes™
towelette and leave it wet for 2 minutes.
CAUTION
To achieve disinfection of the stated efficacy kill claim, the reusable lancing
device should remain visibly wet for 2 minutes right after the completion of
disinfecting treatment.
5. Dispose of the used wipes.
6. Wash your hands thoroughly with soap and clean water.