MH EDS O2D2 Скачать руководство пользователя страница 2

2

BASIC SAFETY

INTRODUCTION

The patented MH EDS-O2D2 is a one or two-user aviation oxygen delivery device. It is designed to deliver aviation oxygen in 
the most efficient, comfortable and convenient way possible.  With its user-selectable settings, apnea alarm and small size, 
the MH EDS-O2D2 is the most portable and flexible electronic digital oxygen delivery system in the world. 

By providing a measured pulse of oxygen every time you breathe in, the MH EDS-O2D2 supplies the oxygen you need to stay 
alert and comfortable while flying. In contrast to constant flow systems that waste oxygen by supplying more than your body 
can use, the MH EDS-O2D2 provides a short pulse of oxygen as you inhale, ensuring that your oxygen is used most efficiently. 
Efficient oxygen delivery means you can fly further on a single oxygen refill or save space and weight with a smaller tank. It 
also makes it more feasible to enjoy the advantages of oxygen while flying below the altitudes where oxygen is mandated--
that can mean fewer headaches, increased alertness, and feeling less exhausted when you reach your destination. Your actual 
oxygen usage will be determined by your breathing rate and physiological needs at altitude.

The programmability of the MH EDS-O2D2 means that, unlike constant-flow oxygen systems, you can “set it and forget it”. 
By automatically detecting your pressure altitude, the MH EDS-O2D2 can be set to start providing oxygen immediately or at 
a specified altitude and will automatically adjust the oxygen flow as your altitude changes.  

When you’re flying, don’t you 

have more important things to do than adjust your oxygen flow during altitude changes?

PILOT

 

#1 Blue 

6 mm 

 

outlet connects to 

Pilot cannula or face mask

Audio port. Sound from

flow-fault and apnea 

alarm conditions only.

(Page 8)

(on bottom)

Static ports, 

DO NOT 

COVER

FEATURES

External 4.5 VDC Power

Connector (Page 8)

(on bottom)

 

•  Easy-to-use two-button control, small size and light weight

 

•  Automatically adjusts oxygen flow with altitude

 

•  Provides reduced oxygen consumption through more efficient oxygen delivery than standard constant-flow systems

 

•  Push button control switch allows automatic altitude enable, Night and Day operations and high flow settings

 

•  Green/Red LED’s indicate oxygen flow, alarm, and status
• Audible and visible apnea alarm informs user of kinked, pinched, or disconnected oxygen lines, obstructed cannula or mask

Red 

4 mm inlet 

connects to the 

regulator

Power 

On/Off

 and 

Mode Switch

#2 Blue 

6 mm

 

outlet

 connects to Co-Pilot

cannula or face mask

Red/Green LED

 

indicator for 

User 1 (Pilot)

Red/Green 

LED

 indicator 

for User 2

Static ports, 

DO NOT 

COVER

• Reduced dry mouth and nose discomfort compared to constant-flow oxygen systems

Pure oxygen is a highly oxidizing gas and can vigorously accelerate combustion.  It can provide a catalyst for spontaneous combus-

tion resulting in personal injury or death if not used properly and with caution.  

DO NOT use any type of oil or grease on 

any of the fittings, valves or cylinders.  DO NOT smoke while in use. DO NOT operate near an open flame.          

DO NOT use the MH EPS (External Power Supply) with the O2D2 as the voltages are not compatible.

2

D-rings

Содержание EDS O2D2

Страница 1: ...ENTS INTRODUCTION 2 FEATURES 2 BASIC SAFETY 2 GETTING STARTED 3 INSERTING OR REMOVING THE BATTERIES 4 REMOVING THE TUBING 4 STORING THE MH EDS O2D2 4 MOUNTING THE MH EDS O2D2 4 THANK YOU for purchasin...

Страница 2: ...omatically adjust the oxygen flow as your altitude changes When you re flying don t you have more important things to do than adjust your oxygen flow during altitude changes PILOT 1 Blue 6 mm outlet c...

Страница 3: ...e next page for detailed instructions 5 If you are using the MH EDS O2D2 with a Mountain High four port regulator FPR locate the oxygen input tube clear tube with a short red tube on one end and inser...

Страница 4: ...be part of your pre flight inventory DO NOT store the EDS unit while the inlet is under pressure Remove all sources of oxygen pressure and secure the unit to ensure it will not become damaged If the l...

Страница 5: ...ment the standard oxygen flow by adding approximately the selected number of feet in thousands to the MH EDS O2D2 s perceived altitude Flow start All altitudes Use with Cannula or face mask Flow amoun...

Страница 6: ...io beeps once every four 4 seconds until unit detects a valid inhalation event Time before Apnea event is 30 35 seconds This typically occurs for the following reasons 1 The user has quit breathing fo...

Страница 7: ...no oxygen pulse on inhalation 1 Make sure the face mask is sealed against the skin 2 Check for obstructions on the O2D2 outlet tubing 3 Use only face masks provided by Mountain High Equipment Supply N...

Страница 8: ...onnect with series diode is provided No internal regulation or over voltage protection is provided External Audio output 3 5 mm stereo type jack provides audio output at 40 mv RMS into 5K L R independ...

Отзывы: