TO INSERT THE TUBING

, push tubing into collar until you

feel  resistance,  Then  push  in  again  approximately  1/4  inch 
farther.   

TO  REMOVE THE TUBING

,  push  in  slightly  on  the  tubing, 

then push in the connector collar while you pull gently on the 
tubing to remove it. 

DO NOT pull on or remove the tubing without 

pushing in the collar; it will damage the connector.

When not being used, the MH EDS-O2D1 unit, oxygen tubes, cannulas, etc., should be disconnected from the oxygen supply and 
stored in a secure manner to ensure that dirt and debris do not enter the inlet and outlet tubes.  The supplied tote bag or a zip-top 
plastic bag is a good storage container.  

If the unit is not going to be used for 30 days or more, remove the 

batteries.  When using the unit for the first time after storage, check the batteries to ensure proper 
operation.  A set of fresh spare batteries should be part of your pre-flight inventory.

 

DO NOT store the EDS unit while the inlet is under pressure.

 Remove all sources of oxygen pressure and secure the 

unit to ensure it will not become damaged. If the lines are disconnected they must be covered so that debris, dust or dirt can’t get 
in. If the supply line is left hooked to the system, make sure that it is first purged with clean dry air or oxygen before the EDS unit is 
connected. 

If the lines are disconnected they must be covered so that debris, dust or dirt can’t get in.

MOUNTING THE MH EDS-O2D1

You may mount the MH EDS-O2D1 unit to a suitable place using the supplied piece of 3M DUAL LOCK tape.  Peel the protective 
backing off one of the rectangles to expose the adhesive and apply it to the back of the unit above the thumb indent for the bat

-

tery door. DO NOT COVER ANY PART OF THE BATTERY DOOR. When a suitable place to mount the unit has been found, peel off the 
protective adhesive backing on the other rectangle and press the adhesive side to the chosen mounting area. 

Let the adhesive 

cure for 12 hours before attempting to remove the two halves of the Dual-Lock from each other, or 
the adhesive may pull away from the unit or mounting area.

INSERTING AND REMOVING THE BATTERIES

STORING THE MH EDS-O2D1

1. Push in the collar

2. Pull the tube  

    straight back 
    while pushing 

    in the collar.

INSERTING AND REMOVING THE TUBING

Remove the battery door by pressing down gently on the battery cover flange 
then slide the door out and away from the unit. The O2D1 unit uses two (2) 
standard  AA  alkaline  batteries.  Insert  the  batteries  as  shown  on  the  label 
inside the battery compartment (they will be a tight fit), then replace the door 
by sliding the door into place until it snaps in place. Take care when removing 
and replacing the batteries as to not damage the batteries and/or connectors.  

Recommended Batteries:

 2 each 1.5 volt DURACELL ULTRA alkaline 

batteries type ‘AA’ or equal quality equivalent.

Battery Life:

 100 Hrs. @ ~25° C. @ ~25% R.H. Measured from 

mean, assuming fresh DURACELL ULTRA alkaline batteries operating under 
normal operating conditions. 

Batteries should be replaced annually or when voltage is low.

Содержание EDS O2D1

Страница 1: ...TURES BASIC SAFETY GETTING STARTED MODE CONTROL SWITCH SETTINGS AND MODES OF OPERATION INSERTING OR REMOVING THE BATTERIES 4 2 2 2 3 TROUBLE SHOOTING WARRANTY SPECIFICATIONS REMOVING THE TUBING STORIN...

Страница 2: ...it By automatically detecting your pressure altitude the MH EDS O2D1 can be set to start providing oxygen immediately or at a specified altitude and will automatically adjust the oxygen flow as your...

Страница 3: ...on the back of the MH EDS O2D1 unit install the 2 AA batteries supplied and replace the battery cover NOTE Batteries fit tightly handle with care See next page for detailed instructions 5 If you are...

Страница 4: ...S O2D1 You may mount the MH EDS O2D1 unit to a suitable place using the supplied piece of 3M DUAL LOCK tape Peel the protective backing off one of the rectangles to expose the adhesive and apply it to...

Страница 5: ...e F10 setting you will receive the effective flow rate of 5 000 10 000 15 000 ft The F modes are useful for people for whom the standard oxygen supply does not achieve the desired blood oxygen saturat...

Страница 6: ...imes with audio beeps once every four 4 seconds until unit detects a valid inhalation event Time before Apnea event is 30 35 seconds This typically occurs for the following reasons 1 The user has quit...

Страница 7: ...aratus SCUBA Self Contained Underwater Breathing Apparatus or any medical operations Before it is put to use it is the responsibility of any user who will use this device to become familiar with the o...

Страница 8: ...and removed during long term storage Nominal battery voltage 2 875 VDC 40 mv 2 25 ma Idle 100 ma Peak 50 ms max 3 25 ma Average Min start up voltage 2 5 VDC 40 mv Low battery cut out 2 0 VDC 40 mv Red...

Отзывы: