Mettler Electronics Corp. — Rev.G_06/28/12
34
Maximum output power:
22 W with a 10 cm² applicator, (ME 7410)
11 W with 5 cm² applicator, (ME 7413)
2.2 W with a 1 cm² applicator (ME 7431)
Maximum intensity:
2.2 W/cm²
Indication accuracy:
±20% (for any level above 10% of maximum)
Output description:
The output waveform is continuous or pulsed as programmed by the
membrane panel control. In the pulse mode the 1.0, 3.2 or 3.3 MHz is pulse
modulated. The power level is adjusted by varying the pulse amplitude.
The
pulse waveform is shown below:
10%
1 ms
9 ms
1 ms
Time
Output
Pulse Space = 10 ms
1, 3.2 or 3.3 MHz
Figure 9.1—10% Pulse Waveform
20%
2 ms
8 ms
2 ms
Time
Output
Pulse Space = 10 ms
1, 3.2 or 3.3 MHz
Figure 9.2—20% Pulse Waveform
50%
5 ms
5 ms
5 ms
Time
Output
Pulse Space = 10 ms
1, 3.2 or 3.3 MHz
Figure 9.3—50% Pulse Waveform