METROTEC
32
11 Technical data
Measurement range
100 % to 10
-31
bar O
2
Ambient temperature
0 to 45 degrees Celsius
Measurement accuracy
+/- 0.3 mV of the sensor EMF
+/- 2 degrees Celsius
+/- 2% of the mA output
+/- 2% of the log oxygen partial pressure
Temperature input
Thermocouple Type S
Heating-up time for sensors
10 to 15 minutes
Response time
approx 2 seconds
Contact load capacity
2A, 24 V (ohmic)
Dimensions
160 x 260 x 260 mm (HxWxD)
Electromagnetic compatibility
The equipment meets the requirements of Euro-
pean directive 89/336EWG and complies with the
following standards:
Interference transmission EN 61000-6-2
Immunity from interference EN 61000-6-4
The device can be used without restriction in res-
idential and industrial surroundings.
Power supply
230 Volt AC
Heating power
approx.. 200 VA
Control power
approx. 40 VA
Temperature of the measurement cell
700
o
C
Temperature detection element
Thermocouple Pt 10Rh-Pt
Temperature of measurement gas
max. 50
o
C
Flow of measurement gas required
At least 8 litre per hour with bypass closed
Moisture
Dew point must be filtered out.
It is important to avoid presence of condensate
Dust
Dust must be filtered out
Solid particles exceeding 25
m
must be filtered
out
Содержание GSM
Страница 10: ...METROTEC 10 4 3 Sketch of the ducting system ...
Страница 15: ...METROTEC 15 ...
Страница 16: ...METROTEC 16 ...