The calibration is necessary to compensate the
cables connected on the DUT, this moves the
reference plane to the end of the cable.
Proceed with calibration by connecting in
sequence SHORT/LOAD(50ohm)/OPEN tool and
then SAVE calibration.
SHORT
LOAD
OPEN
SHORT
LOAD
OPEN
Reference plane
FX700/FX1300 offers two types of calibration:
1) Hardware calibration*
2) OSL user calibration.
The HW calibration is made in the factory. User
can redo it by
positioning the rear switch to the
right
position and proceeding with the
calibration. This calibration is done only once
and it is not necessary to repeat it. (The switch
could be inside the enclosure, in this case you
need to remove the rear cove).
After calibration,
reposition the rear switch to the left position.
OSL kit* not included
*use a better quality calibration kit.
Reference plane