VERSION

PAN_F1

14000

Panoramic window initial frequency, kHz

PAN_SPAN

2

Span for panoramic window (0..9)

MEAS_F

14000000

Measurement window frequency, Hz

SYNTH_TYPE

SI5351A

Frequency synthesizer type. 0 for Si5351, do not change: other
values are reserved for future use.

SI5351_XTAL_FREQ

27000000

Si5351a

s nominal crystal frequency, Hz

SI5351_BUS_BASE_

C0h

Si5351a address on I2C bus. Can be changed to any even number in

ADDR

the range 02h..FEH

SI5351_CORR

0

Si5351a

s crystal frequency offset from nominal frequency, Hz

OSL_SELECTED

A

Selected OSL file index (0..15 for files A .. P, other values - no OSL
file selected)

Z0

50

Base Z0 for VSWR measurements and Gamma calculation

OSL_RLOAD

50

OSL calibration standard for the Load measurement, Ohm

OSL_RSHORT

0

OSL calibration standard for the Short measurement, Ohm

ROPEN

open

OSL calibration standard for the Open measurement, Ohm. Use
value >100000 for purely open load.

OSL_NSCANS

1

Number of scans to average during OSL calibration

MEAS_NSCANS

1

Number of scans to average in measurement window

PAN_NSCANS

1

Number of scans to average in panoramic window

LIN_ATTENUATION

6 (06h)

Linear audio input attenuation. Applied during audio input device
initialization after reset. Sets the linear audio input volume to (100 -
ATTENUATION). One unit is approximately 1 dB.

F_LO_DIV_BY_TWO

0

Deprecated, references in the code have been removed. Leave at
default setting (0). Set to non zero if LO frequency is divided by two
in quadrature mixer. Intended to be used with obsolete version 2 RF
front end.

GEN_F

14000000

Frequency for generator window, Hz

PAN_CENTER_F

0

Way of setting panoramic window. 0: F0+band_span, 1: F_/-
Band_span/2

BRIDGE_RM

5.1

Value of measurement resistor (R4, R11) in bridge, Ohm float

BRIDGE_RADD

200

Value of series resistor (R5, R9) in bridge, Ohm

BRIDGE_RLOAD

51

Value of load resistor in bridge (R7, R10), Ohm

COM_PORT

USB

Serial (COM) port to be used: 0 for COM1, 1 for COM2

COM_SPEED

38400

Serial (COM) port speed, bps

LOWPWR_TIME

off

Time in milliseconds after which to lower power consumption by
switching off LCD. (0 - disabled)

* the information in this table may vary with an updated firmware version

Содержание FX1300

Страница 1: ...1 Ver 1 0 5 www metropwr com ...

Страница 2: ... malfunction turn it off immediately IMPORTANT INFORMATION Read all the operating instructions safety tips and warnings in the instruction manual Identifying potential hazardous situations and observing the appropriate safety rules will avoid accidents Dangerous situations to avoid in order to prevent all risks that are shown above Never use the FX700 inappropriately but only as described in the u...

Страница 3: ...can be used Fast check of all antenna parameters Coaxial cable length measurement Export graphics to your computer Detailed screenshots of the measures Quickly measure resonance of multiband antennas Comparison of many types of antennas Generate RF signals Measurement of the series and parallel parameters of the quartz Complete OSL calibration open load short TDR chart time domain reflectometer It...

Страница 4: ...the antenna to the DUT short the antenna connector to ground Up Down change frequency don t remove it Sweep Span Enter manual Frequency OM Band USB PC PORT DUT N type Command POWER ON OFF SD Card USB FOR BATTERY CHARGER Display 4 3 ...

Страница 5: ...5 Marker Sweep scan FAST ACCURATE Move marker right FREQ Z SWR MCL Q Zoom SWR Tap here change graph type SAVE Screenshot SWR R X graph Smith chart graph Grap S11 Returnloss Tap here enter frequency ...

Страница 6: ...nds This function is very useful for measuring multiband antennas Tap here to enter frequency frequency SWR measure Tone disable Tap here to change the type of series parallel measurement Antenna equivalent circuit Easy Menu enter the antenna frequency and a modulated tone warns when SWR is lower ...

Страница 7: ... basic parameters of quartz Fs Fp resonance serial parallel Cs Cp capacity serial parallel Ls inductance Rs resistance Q quality factor Menu Generator possibility to select the frequency and set steps A square wave and all harmonic frequencies will be generated on the DUT connector no filtering expected Equivalent circuit of a quartz KEYB set frequency quartz CAL OPEN disconnect cable on N connect...

Страница 8: ... offers two types of calibration 1 Hardware calibration 2 OSL user calibration The HW calibration is made in the factory User can redo it by positioning the rear switch to the right position and proceeding with the calibration This calibration is done only once and it is not necessary to repeat it The switch could be inside the enclosure in this case you need to remove the rear cove After calibrat...

Страница 9: ...trace size of graph menu BEEP ON OFF active deactivate beep This is a service menu that displays the spectrum of V I voltage current signals taken from the measurement coupler Through this menu it is possible to select and view the screenshots stored on the sd card In this area it is possible to manage the screenshot files ...

Страница 10: ...u 0 hide SCREENSHOT_FOR 0 Screen shot image file format 1 PNG 0 BMP MAT BAND_FMAX 710MHz Maximum operating frequency of the device from 150 to 710MHz SI5351_MAX_FREQ 200MHz Maximum frequency in Hz that Si5351a can reliably output Can be 160MHz or 200 MHz SI5351_CAPS 10pF Si5351 built in quartz capacitors setting 1 for 6pF 2 for 8pF 3 for 10 pF TDR_VF 66 Velocity factor for TDR distance calculation...

Страница 11: ...ring OSL calibration MEAS_NSCANS 1 Number of scans to average in measurement window PAN_NSCANS 1 Number of scans to average in panoramic window LIN_ATTENUATION 6 06h Linear audio input attenuation Applied during audio input device initialization after reset Sets the linear audio input volume to 100 ATTENUATION One unit is approximately 1 dB F_LO_DIV_BY_TWO 0 Deprecated references in the code have ...

Страница 12: ... can be interfaced with the PC via the left USB port At the moment it is compatible with third party software Through the second right USB port it is possible to download screenshots on the computer in bmp png format ...

Страница 13: ...13 USB port Drag drop files in the windows Reboot FX700 FX1300 Connect USB cable ...

Страница 14: ...l Smith chart measure antenna multiband TDR time domain reflectometer Cable length measurement Storage of graphics on the sd bmp png card Export images via sd Measure quartz parameters Menu generator Fast accurate scan Battery LI ion included 2600mA USB charging circuit Firmware upgradable via USB 3 7V operating voltage Compact dimensions 135 x 32 x 85 mm Weight 250gr www metropwr com Technical fe...

Отзывы: