DL2
70
13.6.2 Totalizer archive
Organization of additional header in the totalizers archive file:
CHANNEL, DESCRIPTION, TOTALIZER 1 TYPE, TOTALIZER 1 UNIT, TOTALIZER 2
TYPE, TOTALIZER 2 UNIT
CHANNEL
– the channel number
DESCRIPTION
– channel description, depending on the settings in the device
TOTALIZER 1 TYPE/
TOTALIZER 2 TYPE
– totalizer operation mode: ‘ ‘ – disabled; ‘1’ – unresettable;
‘2’ – resettable; ‘3’ – daily; ‘4’ – weekly; ‘5’ – monthly
TOTALIZER 1 UNIT/
TOTALIZER 2 UNIT
– unit assigned to the totalizer
Organization of records in the totalizers archive file:
DATE, TIME, DST, CH1:T1, CH1:T2, CH12:T1,
… , CH22:T2, CRC2
DATE
– date of the record in YY-MM-DD format
TIME
– time of the record in HH:MM:SS format
DST
– information about summer (1) or winter (0) time enabled
CHXX:TY
– value of totalizer (XX – channel number from 1 to 30; Y – totalizer
number: 1 or 2)
CRC2
– CRC control
13.6.3 Event archive
Organization of records in the events archive file:
DATE, TIME, DST, EVENT CODE, CRC2
DATE
– date of the record in YY-MM-DD format
TIME
– time of the record in HH:MM:SS format
DST
– information about summer (1) or winter (0) time enabled
EVENT CODE
– event code (more information below)
CRC2
– CRC control
The event archive records the following events:
SYS:STOP
Power supply switched off
SYS:START
Power supply switched on
ARCH:NEW
Created a new archive file
ARCH:STOP
Stopped archivization
ARCH:START
Started archivization
SYS:NEW SETTINGS
New settings saved
SYS:TIME CHANGED
Changed time
SYS:DATE CHANGED
Changed date
SYS:CHx: AUX VALUES RESET
Reset auxiliary values for "x" channel
(min, max, totalizers)