Resolution:
1 V L-N
2.
Current Capacity
15 Amps @ 100 Volts
15 Amps @ 120 Volts
10 Amps @ 230 Volts
3.
Resistance Measurement
DC-Source:
Range:
0 - 19.99 Ohms
Accuracy:
± 2%reading, ± 1 digit
Resolution:
0.01 ohms
Current Source:
100 mA DC
AC-Source:
Range:
0 - 1999 mOhms
Accuracy:
± 2% reading, ± 1 digit
Resolution:
1 mOhm
Current Source:
200 mA AC
4.
Leakage Current
Range:
0 - 1999 µA RMS
Accuracy:
DC and 25 to 1.0 kHz, ± 2% reading, ± 1 digit
1.0 kHz to 100 kHz, ± 2.5%R, ± 1 digit
100 kHz to 1 MHz, ± 5.0%R, ± 1 digit
Resolution:
1 µA
5.
Input Impedance
1000 ohm, AAMI ES1-1993 load or
IEC 60601.1 load
6.
Power Consumption
110 V (50/60) Hz = 200 mA
115 V (50/60) Hz = 200 mA
240 V (50/60) Hz = 100 mA
2
Содержание QA-ST
Страница 1: ...QA ST User and Service Manual QA ST Safety Tester North American Version 1 1...
Страница 4: ...This page intentionally left blank iv...
Страница 6: ...This page intentionally left blank vi...
Страница 13: ...This page intentionally left blank 3...
Страница 14: ...3 Operating the QA ST This chapter describes the controls and terminals of the QA ST Safety Tester 1...
Страница 20: ...4...
Страница 27: ...2 Press Patient leakage current lead to GND Ground 3 The value will be displayed on the LCD Note the value 4 11...
Страница 29: ...13...
Страница 34: ...7 Perform the test for tall the given values in the test protocol un der Ground Wire Resistance DC Reference 5...
Страница 43: ...8 Perform the test for tall the given values in the test protocol un der Leakage Current to ISO Reference 8 14...
Страница 44: ...Appendix A Circuit Board Schematic North American Version A 1 A 1...
Страница 45: ...This page intentionally left blank A 2...