10. Perform the test for tall the given values in the test protocol
under “Leakage Current Int. Case: Reference”.
5.5 Le
akage
Cur-
rent –
Exter-
nal
Case
The purpose of this test is to determine if there is any leakage cur-
rent between DUT Enclosure (chassis) and Mains Earth.
6.
Connect a voltage source in series with an ammeter and
10K
Ω
resistance as shown below.
7.
Press “Leakage current Ext. Case”.
8.
Adjust the voltage until the ammeter displays the first refer-
ence value in the test protocol. QA-ST will display the same
value as the reference ammeter. Note the value.
9.
Next, press “Outlet Supply RS” (Reversed Supply). The dis-
played QA-ST value will not change.
7
Содержание QA-ST
Страница 1: ...QA ST User and Service Manual QA ST Safety Tester North American Version 1 1...
Страница 4: ...This page intentionally left blank iv...
Страница 6: ...This page intentionally left blank vi...
Страница 13: ...This page intentionally left blank 3...
Страница 14: ...3 Operating the QA ST This chapter describes the controls and terminals of the QA ST Safety Tester 1...
Страница 20: ...4...
Страница 27: ...2 Press Patient leakage current lead to GND Ground 3 The value will be displayed on the LCD Note the value 4 11...
Страница 29: ...13...
Страница 34: ...7 Perform the test for tall the given values in the test protocol un der Ground Wire Resistance DC Reference 5...
Страница 43: ...8 Perform the test for tall the given values in the test protocol un der Leakage Current to ISO Reference 8 14...
Страница 44: ...Appendix A Circuit Board Schematic North American Version A 1 A 1...