5-6
e.
Use a current source to generate an AC current between Pa-
tient Leads 1-10 and EARTH. Check that both the level of
this current, and the current readout on the QA-90 display,
are equal. Press
STOP (F1)
.
f.
Press the
PL
key on the keypad, then , press
PREV (F3)
or
NEXT (F4)
to ensure that the module with Patient Lead 11
(blue lead) is active.
g.
Press
GO BACK (F2)
to return to the test window, then
press
START (F1)
and run the same test between Patient
Lead 11 and EARTH. Repeat the test with different current
values. Press
STOP (F1)
.
h.
Press
GO BACK (F2)
to return to Manual Test Setup Win-
dow 3.
4.
Patient Leakage Current DC
a.
Press
Patient Leakage Current DC (F5)
.
b.
Select
Normal Condition (F7)
.
c.
Press the
PL
key on the keypad to ensure that the module
with Patient Leads 1-10 (white leads) is active. If not, press
PREV (F3)
or
NEXT (F4)
so that it is active.
d.
Press
GO BACK (F2)
to return to the test window, then
press
START (F1)
.
e.
Use a current source to generate an DC current between Pa-
tient Leads 1-10 and EARTH. Check that both the level of
this current, and the current readout on the QA-90 display,
are equal. Press
STOP (F1)
.
f.
Press the
PL
key on the keypad, then , press
PREV (F3)
or
NEXT (F4)
to ensure that the module with Patient Lead 11
(blue lead) is active.
g.
Press
GO BACK (F2)
to return to the test window, then
press
START (F1)
and run the same test between Patient
Lead 11 and EARTH Repeat the test with different current
values. Adjust the General I-constant if the values between
UUT and reference are not equal. Press
STOP (F1)
.
h.
Press
GO BACK (F2)
until you return to Manual Test Setup
Window 3. Then, press
MORE (F1)
to advance to Manual
Test Setup Window 4.
Patient Auxiliary Current AC ..........> F7
Patient Auxiliary Current DC ..........> F6
Mains on Applied Part .................> F5
Содержание QA-90
Страница 1: ...QA 90 User Service Manual QA 90 Electrical Safety Analyzer P N 11025 V 3 13...
Страница 14: ...2 4 This page intentionally left blank...
Страница 30: ...3 16 This page intentionally left blank...
Страница 35: ...4 5 Test according to IEC60601 1 F7 Test Type Rapid F6 Test Mode Manual F5 GO BACK START F1 F2 F3 F4...
Страница 40: ...4 10 MANUAL TEST SETUP F7 Mains Voltage F6 Current Consumption F5 MORE MAIN MENU F1 F2 F3 F4...
Страница 44: ...4 14 This page intentionally left blank...
Страница 51: ...5 7 MORE GO BACK MAIN MENU F1 F2 F3 F4...
Страница 60: ......
Страница 68: ...6 8 This page intentionally left blank...
Страница 86: ...A 18 This page intentionally left blank...
Страница 88: ...B 2 This page intentionally left blank...
Страница 90: ...B 4 Measuring Board Component Location Layer 1...
Страница 91: ...B 5 Measuring Board Component Location Layer 2...
Страница 92: ...B 6 Measuring System Schematic Diagram Part 1 Measuring Preparation...
Страница 93: ...B 7 Measuring System Schematic Diagram Part 2 High Voltage Interface...
Страница 94: ...B 8 Measuring System Schematic Diagram Part 3 High Voltage Logic...
Страница 97: ...B 11 Measuring System Schematic Diagram Part 6 Measurement Matrix...
Страница 98: ...B 12 Measuring System Schematic Diagram Part 7 Program Amplifier and Lowpass...
Страница 99: ...B 13 Measuring System Schematic Diagram Part 8 Relay Drivers...
Страница 100: ...B 14 Microprocessor Board Component Location...
Страница 102: ...B 16 Microprocessor System Schematic Diagram Part 1 QA 90...
Страница 103: ...B 17 Microprocessor System Schematic Diagram Part 2 Integrated Keypad...
Страница 104: ...B 18 Microprocessor System Schematic Diagram Part 3 CPU...
Страница 105: ...B 19 Microprocessor System Schematic Diagram Part 4 CPU...
Страница 106: ...B 20 Microprocessor System Schematic Diagram Part 5 Printer and Display Interface...
Страница 108: ...B 22 This page intentionally left blank...
Страница 110: ...C 2 This page intentionally left blank...
Страница 112: ...D 2...