METRON ACCUSONIC PLUS Скачать руководство пользователя страница 6

Effective Radiating Area

Nominally 0.8 square cms +/- 20%

 (1 cm head)

Nominally 5.0 square cms +/- 20% 

(2.5 cm head)

Power/Intensity Display

Accurate to +/- 10% of reading for
outputs in excess of 1 Watt.

ULTRASOUND MODULATION:

Modulation Modes

Continuous
Pulsed

Pulsed Modulation

Pulse Frequency

100 Hz +/- 2%

Pulse Width

1.0, 2.0, 5.0 millis/- 2%

Pulse Duty Cycle

1:9 (10%), 1:4 (20%), 1:1 (50%)

respectively

BEAM NON-UNIFORMITY RATIO:

5:1 +/- 20%                  

TREATMENT TIMER:

Maximum Treatment Time

30 m/- 2%

At Time Expiration

Time display zeros and a 3 second 
audible alarm sounds.

CONTACT CONTROL:

Function

To detect poor acoustic coupling between the 
ultrasonic treatment applicator and the patient.

On detection of contact loss

The LED indicator on the selected applicator 

handle turns red and, after a 2 second delay, the 
treatment timer is halted. The ultrasonic output is 
reduced to 0.2 Watts/cm

2

.

On detection of contact

The treatment timer is restarted and the ultrasonic 
output power is restored to the selected value.  
The LED indicator on the applicator handle turns 
green.

Purpose/Rationale

Ensures the patient receives the required 
ultrasound dose and prevents damage to the 
ultrasonic treatment applicator by heat generated 
in the transducer when it is operated unloaded at 
high power levels.

Metron                                                                                  Accusonic Plus

__________________________________________________________________

   

Version 1.6

March 2002

Page  2  of  12

Содержание ACCUSONIC PLUS

Страница 1: ...57 Aster Avenue Carrum Downs Victoria Australia 3201 Tel 03 9775 1234 Fax 03 9775 1990 Int 61 3 9775 1234 Int 61 3 9775 1990 TECHNICAL MANUAL METRON ACCUSONIC PLUS ULTRASOUND THERAPY UNIT MODEL AP 100...

Страница 2: ...r Metron Accusonic Plus Ultrasound Therapy Unit Model AP 100 is in conformity with the essential requirements and provisions of Council Directive 93 42 EEC is subject to the procedure set out in Annex...

Страница 3: ...Should you as the user of this technical manual wish to make any comment about the product or this manual our Authorised Representative within the European Union may be contacted as follows Metron Med...

Страница 4: ...7 3 2 Ultrasonic Treatment Applicator s 7 4 Preventative Maintenance Quality Assurance 8 4 1 Calibration and Adjustment 8 4 1 1 Equipment Required 8 4 1 2 Calibration Procedure 8 4 2 Electrical Safet...

Страница 5: ...equency 1 1 MHz 10 Output Intensity Continuous Mode 3 0 Watts cm2 20 Maximum Power Intensity Display Accurate to 20 of reading for outputs in excess of 0 2 Watts cm2 Effective Radiating Area 5 0 squar...

Страница 6: ...ling between the ultrasonic treatment applicator and the patient On detection of contact loss The LED indicator on the selected applicator handle turns red and after a 2 second delay the treatment tim...

Страница 7: ...pment IEC 601 1 1988 Medical electrical equipment Part 1 General requirements IEC 601 2 5 1984 Medical electrical equipment Part 2 Particular requirements for the safety of ultrasonic therapy equipmen...

Страница 8: ...il or fax as detailed below Metron Medical Australia Pty Ltd P O Box 2164 CARRUM DOWNS AUSTRALIA 3201 Fax 03 9775 1990 from within Australia 61 3 9775 1990 International The Accusonic Plus generates c...

Страница 9: ...er is reduced from that selected to 0 2 Watts cm2 At detection of contact the treatment timer resumes and the ultrasonic output power is restored to the selected value This ensures the patient receive...

Страница 10: ...icrocontroller and the tune point of the applicators is stored in the eeprom located in the connector of the treatment applicator L5 monitors supply current level in the output stage Comparing this va...

Страница 11: ...ive RF from turning it on The piezoelectric resonator which generates the ultrasound energy is permanently bonded to the cap of the applicator It can be replaced by simply unscrewing the cup and reson...

Страница 12: ...display will show a P indicating the next step the alignment of the applicator s output power and contact sense control is ready to be undertaken Place the applicator in the power meter Press the Mode...

Страница 13: ...confirm continued operator and patient safety Other mandatory statutory requirements for electrical safety inspections may also apply 4 3 Disassembling Assembling the Unit If required disassembly and...

Страница 14: ...5 SCHEMATIC DIAGRAMS 5 1 Main Circuit Board Layout Metron Accusonic Plus __________________________________________________________________ Version 1 6 March 2002 Page 10 of 12...

Страница 15: ...5 2 Schematic Diagram Main Printed Circuit Board Metron Accusonic Plus __________________________________________________________________ Version 1 6 March 2002 Page 11 of 12...

Страница 16: ...5 3 Schematic Diagram Treatment Head Printed Circuit Board Metron Accusonic Plus __________________________________________________________________ Version 1 6 March 2002 Page 12 of 12...

Отзывы: