FDI2056 Use
r’s Manual
v 2.0 r 1.1
– 03/13
Copyright © 2013 Metrolab Technology SA
75
www.metrolab.com
Acquisition and Trigger must be started before any triggers can be sent. It means also that the Self
Trigger cannot be started simultaneously with the two others. Otherwise, the first partial integral might be
wrong. Doing this initialization by writing two successive values will guarantee enough time for the
system to settle.
Interrupt Enable (BAR3 : 0x102)
Bit 0
Enable status bit 0 condition
Bit 1
Enable status bit 1 condition
Bit 2
Enable status bit 2 condition
Bit 3
Enable status bit 3 condition
Bit 4
Enable status bit 4 condition
Bit 5
Enable status bit 5 condition
Bit 6
Enable status bit 6 condition
Bit 7
Global interrupt enable
Interrupt Pending (BAR3 : 0x103)
Bit 0
Status bit 0 condition is pending
Bit 1
Status bit 1 condition is pending
Bit 2
Status bit 2 condition is pending
Bit 3
Status bit 3 condition is pending
Bit 4
Status bit 4 condition is pending
Bit 5
Status bit 5 condition is pending
Bit 6
Status bit 6 condition is pending
Bit 7
unused
Gain configuration (BAR3 : 0x104)
Register value
Gain
0
0.1
1
0.2
2
0.4
3
0.5
4
1.0
5
2.0
6
4.0
7
5.0
8
10.0