FDI2056 Use
r’s Manual
v 2.0 r 1.1
– 03/13
Copyright © 2013 Metrolab Technology SA
54
www.metrolab.com
010 Timer with Synchronization
TRS,T,S
011 Encoder without index nor synchronization
TRS,E
100 Encoder with external synchronization
Linear with Index
TRS,E,S
101 Rotating Encoder with index
TRS,E,dddd
110 External Trigger
TRS,X
111 External Trigger with synchronization
TRS,X,S
Bit 4
Infinite sequence
This bit is set to indicate that an infinite trigger sequence is in progress.
An infinite sequence is set using ni=* in the
TRI
command.
Bit 3
Measurement in progress
This bit is set to indicate that the FDI 2056 is in the process of
measuring. The bit is automatically set by the
RUN
command; it is
automatically cleared when the sequence is terminated.
Bit 2
Forwards / Backwards
A value of 1 indicates motion in the forward direction (FW LED lit), a
value of 0 refers to the backward direction (BW LED lit).
Note that when no encoder is in use, this bit has no meaning.
Bit 1
Motor +
This bit indicates that the motor has been activated in the positive
direction (corresponds to FW if the cabling is correct), see the
MOT,+
command for more details.
This bit is set or cleared even if the software motor interface is unused;
see chapter 4-3 for more details.
Bit 0
Motor -
This bit indicates that the motor has been activated in the negative
direction (corresponds to BW if the cabling is correct), see the
MOT,-
command for more details.
This bit is set or cleared even if the software motor interface is unused;
see section 4-3 for more details.