66
▪▪▪▪▪▪▪
7.
With the lamp wires straight, lift the lamp
retaining ring upward and off the lamp as
shown.
Discard this lamp immediately. Order a
new lamp from your Metrohm distributor
to keep for the next lamp change.
Always keep a spare lamp in stock, to
avoid last-minute emergencies. The part
number is XA-3000.
8.
Unpack the new lamp from the protective
box. Place the new lamp carefully onto
the bench, and place the lamp retaining
ring over the wires as shown.
9.
Gently place the lamp and retaining ring
into the lamp box in the instrument.
The outer edge of the lamp should fit into
the machined inset in the lamp plate.