4.5 Pump tubings
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
919 IC Autosampler plus
Pump tubing selection
Pump tubing can differ in terms of material, diameter and thus flow rate.
Different pump tubing is used depending on the application.
The following table shows the properties and areas of application of the
pump tubings:
Table 2
Pump tubings
Order
number
Name
Material
Inner
diameter
Use
6.1826.020
Pump tubing (blue/
blue), 2 stoppers
PVC
(Tygon
®
ST)
1.65 mm
Pump tubing for online IC
instruments and automa-
tion in voltammetry.
6.1826.310
Pump tubing LFL
(orange/green), 3
stoppers
PVC
(Tygon
®
)
0.38 mm
Pump tubing for bromate
determination using the
triiodide method.
6.1826.320
Pump tubing LFL
(orange/yellow), 3
stoppers
PVC
(Tygon
®
)
0.48 mm
For acceptor solutions for
Inline Dialysis and for Inline
Ultrafiltration.
6.1826.330
Pump tubing LFL
(orange/white), 3
stoppers
PVC
(Tygon
®
)
0.64 mm
No special applications.
6.1826.340
Pump tubing LFL
(black/black), 3 stop-
pers
PVC
(Tygon
®
)
0.76 mm
For sample solution in
Inline Dialysis.
6.1826.360
Pump tubing LFL
(white/white), 3 stop-
pers
PVC
(Tygon
®
)
1.02 mm
For sample transfer.
6.1826.380
Pump tubing LFL
(gray/gray), 3 stoppers
PVC
(Tygon
®
)
1.25 mm
For Inline Dilution.
6.1826.390
Pump tubing LFL (yel-
low/yellow), 3 stop-
pers
PVC
(Tygon
®
)
1.37 mm
For sample solution in
Inline Ultrafiltration.