Insert new module for the Autolab 8-Series
22 |
P a g e
Figure 21 – Verify that the value of C1 and C2 are set to 0.0 before starting the calibration
13.1
– Determination of the C1 and C2 parameter in Nova
The value of the C1 and C2 parameters must be measured and modified in the
FRA2CAL.INI file (for the FRA2 module only). The determination of C1 and C2 requires
the following items:
•
Autolab Dummy cell
•
Faraday cage
13.1.1
– Determination of C1
Follow the steps described in this section to determine the value of the C1 parameter.
1.
Start the Nova software.
2.
Be sure C1 and C2 are set initially to 0.0 (see previous section and Figure 21).
3.
Import the procedure
PGSTAT C1 calibration.nox from the NOVA installation
folder
3
The default installation folder is C:\Program Files\Metrohm Autolab\SharedDatabases\Module test.