Insert new module for the Autolab 8-Series
16 |
P a g e
10
– Analog bus adjustments for the BA module
For the BA module the SMB coaxial cables should be connected between the BA module
and the PGSTAT and the FRA2 or FRA32M, if applicable. Note: The supplied jumper cable
is NOT used for the BA module in an 8-Series instrument!
10.1
– If the FRA2 module is present
For the BA module the SMB coaxial cables should be connected as shown in Figure 15.
Figure 15 – Overview of the SMB cables required between the BA module and the PGSTAT and
optional FRA2 module
4.
Disconnect the three SMB cables between the FRA2 and the PGSTAT and
reconnect these cable to the BA module:
a.
Connect DSG signal of the FRA2 to FRA DSG of the BA.
b.
Connect Ein of the FRA2 to E-FRA of the BA.
c.
Connect Iin of the FRA2 to I-FRA of the BA.
5.
Connect the special copper labelled SMB cable between the WEx of the BA
module and the BIPOT WE2 of the PGSTAT. Please note that the location of BIPOT
WE2 on the PGSTAT differs slightly depending on the model (see Figure 15).
6.
Connect the three additional SMB cables:
a.
Connect To PGstat of the BA to DSG of the PGSTAT.
b.
Connect E-PGSTAT of the BA to FRA E of the PGSTAT.