3.3 Fitting the accessories
838 Advanced Sample Processor
, Installation
15
Please refer to the 786 Swing Head Instructions for Use, which you can
obtain from Metrohm by quoting order No. 8.786.1001, for details of the
individual settings and robotic arm types.
3.3.3 Fitting the retention plate
When sampling from covered vials with the needle, the retention plate
9
serves to wipe the vials when the lift is rising. It is fixed in position on the
front side of the tower with the three supplied Allen screws.
Figure 13 Fitting the retention plate
3.3.4 Fitting the needle or tubing
22
PEEK needle (6.1835.010)
or
PEEK tubing (6.1835.020 /
6.1835.040 / 6.1835.050)
23
Needle holder
24
PEEK pressure screw (4.766.4320
or
6.2833.020 +
6.2744.080)
25
PEEK pressure screw (6.2744.010)
26
PTFE capillary (6.1803.070
or
6.1831.050 / 6.1831.060 /
6.1831.080)
When using a 6.1835.xxx PEEK tubing,
do not use any
stoppers
on the sample vessels since these cannot be
pierced by the tubing and this would damage the PEEK tub-
ing!
With PEEK needles
perforated stoppers
can be used.
Figure 14 Fitting the needle
22
Ring wedge
23
24
25
9
26