4 Malfunctions - Maintenance
793 IC Sample Prep Module/ 8.793.1003 Instructions for Use
36
7 Connect and condition the SP Module A
x
Reconnect the actuator to the IC system (see
section 2.3.2
and section 2.4.2
).
x
Before switching the actuator to the next position for the first
time rinse all 3 actuator units with solution for 5 min.
8 Check braking range
x
Check the braking range of the SP Module A according to
section 4.2.8
and readjust if necessary.
4.2.8 Braking range adjustment
It may be necessary to readjust the braking range of the 793 IC Sample
Prep Module in the following cases (see also
section 4.1
):
x
Actuator switches to next position but display lamp
7777
no
longer lights up
x
When the actuator is exchanged (see
section 4.2.7
)
The braking range is adjusted by means of adjustment screw
18
18
18
18
"Breakpoint
'
t" on the rear panel of the 793 IC Sample Prep Module.
Proceed as follows:
1 Condition the cation exchange material
x
If the actuator has been cleaned or exchanged then before
switching the actuator to the next position for the first time all
3 actuator units must be rinsed with solution for 5 min.
2 Check the position of the Actuator rotor
x
Switch the instrument off and then on again. Switch off actua-
tor remote operation with key
6666
<REMOTE> and operate key
8888
<STEP>.
x
By rotating the SP Module A or with the aid of a mirror check
the position of the actuator rotor according to
Fig. 10
: at least
one of the three holes of the actuator rotor must be partly or
completely visible through the opening at the bottom of the
actuator block.
x
Repeat the procedure 2-3 times; wait at least 10 s each time
before the <STEP> key is operated.
x
If the rotor stops in an unacceptable position then the braking
range must be readjusted (continue with point
3
).