3.2. Diagnosis
776 Dosimat
28
22.
Depending on the exchange unit used, enter the volume as below:
1 ml: 0.02 ml
5 ml: 0.1 ml
10 ml: 0.2 ml
20 ml: 0.4 ml
50 ml: 1
ml
23.
<ENTER>
24.
<GO> (depress briefly) and with a stop watch take the time until the cock starts
turning.
The time must be 14 ... 24 s.
25.
Reset the parameter if under 5. the reading was auto fill = on .