4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
76
.Cond:
Conditioning ON/OFF
.DCor.Type: Type of drift take-over for the drift correction. auto: Take-over of
the drift value at start.
.DCor.Value: Drift value for the manual drift correction.
4.2.2.15. M
ode
.P
arameter
.P
resel
.I
Req
id1, id1&2, all,
OFF
M
ode
.P
arameter
.P
resel
.S
Req
value
, unit, all,
OFF
M
ode
.P
arameter
.P
resel
.R
eqTitr
ON
,
OFF
Automatic inquiries after the start of the determination. From such an inquiry,
the determination continues if the requested entry/entries is/are made, e.g.
&SmplData.OFFSilo.Id1 (see 4.2.2.56) or with &M $G, see 4.2.2.1.
.ReqTitr: Current generation starts during requests (with ON).
4.2.2.16. M
ode
.P
arameter
.P
resel
.Sa
mpleUnit
g
, ...up to 5 ASCII
Method specific sample unit, i.e. when the method is loaded, the current unit
of the sample size is overwritten by the unit from the method.
4.2.2.17. M
ode
.P
arameter
.P
resel
.L
imSmplSize
.S
tatus ON,
OFF
M
ode
.P
arameter
.P
resel
.L
imSmplSize
.L
oLim
0.0
...999 999
M
ode
.P
arameter
.P
resel
.L
imSmplSize
.U
pLim 0.0...
999 999
Limit control for the sample size.
4.2.2.18. M
ode
.P
arameter
.P
resel
.Id
1Text
id1/C21
, 10 ASCII characters
M
ode
.P
arameter
.P
resel
.Id2
Text
id2/C22
, 10 ASCII characters
M
ode
.P
arameter
.P
resel
.Id3
Text
id3/C23
, 10 ASCII characters
Text for sample identifications.
4.2.2.19. M
ode
.P
arameter
.P
resel
.Ce
ll
no diaph.
, diaphragm
M
ode
.P
arameter
.P
resel
.G
enI 100,
200,
400
, auto
.Cell:
Type of generator electrode.
.GenI:
Current at the generator electrode in mA. "auto" means that the
current is switched in the course of determinations.
Default: 400 mA for cells without diaphragm, auto for cell with
diaphragm.
4.2.2.20. M
ode
.P
arameter
.P
resel
.O
ven COM1,
COM2,
no
If an Oven is connected, its results will be incorporated into the result report of
the Coulometer.
If there is no Oven connected via RS232, this parameter has to be on "no".
4.2.2.21. M
ode
.P
arameter
.P
resel
.A
ctPuls
first, all, cond.,
OFF
Output of a pulse on the I/O line "Activate", see page 132.
4.2.2.22. M
ode
.D
ef
.F
ormulas
.1.F
ormula
H2O, CXX, RSX, +, -, *, /, (, )
M
ode
.D
ef
.F
ormulas
.1.T
extRS
8 ASCII characters