4.1 General rules
756/831 KF Coulometer, Instructions of Use
53
4 Operation via RS232 Interface
4.1 General rules
The KF Coulometer has an extensive remote control facility that allows full
control of the KF Coulometer via the RS 232 interface, i.e. the KF Coulometer
can receive data from an external controller or send data to an external con-
troller. CR and LF are used as terminators for the data transfer. The KF Cou-
lometer sends 2xCR and LF as termination of a data block, to differentiate be-
tween a data line which has CR and LF as terminators. The controller termi-
nates its commands with CR and LF. If more than one command per line is
sent by the controller, “;” is used as a separator between the individual com-
mands.
The data are grouped logically and easy to understand. Thus e.g., for the se-
lection of the dialog language, the following must be sent
&
C
onfig.
A
ux.
L
anguage
"english"
whereby it is sufficient to only transmit the boldface characters, thus:
&
C
.
A
.
L
"english"
The quantities of the commands above are:
C
onfig configuration
data
A
ux
auxiliaries, various data
L
anguage
setting the dialog language
The data are hierarchically structured (tree form). The quantities that occur in
this tree are called
objects
in the following
.
The dialog language is an object
which can be called up with the
&
C
onfig.
A
ux.
L
anguage
command.
If one is in the desired location in the tree, the value of the object can be que-
ried.
&
C
onfig.
A
ux.
L
anguage
$Q
Q means Query
The query command $Q initiates the issuing of the value on the instrument
and the value emission is triggered. Entries which start with $, trigger some-
thing. They are thus called
triggers.
Values of objects can not only be queried, they can also be modified. Values
are always entered in quotes, for example:
&
C
onfig.
A
ux.
L
anguage
"english"