Part# 100951
• REV A
(January 2018) Page
5 of 5
www.metrixvibration.com
8824 Fallbrook Dr. Houston, TX 77064, USA
Tel: 1.281.940.1802 • Fax: 1.713.559.9421
After Hours (CST) Technical Assistance: 1.713.452.9703
7.
Slide the connector onto the center core pin and ferrule. Be careful to ensure the
center core pin comes through the connector. Crimp the connector, using the .178
hexagon hole, at the ridge line. Test the connection by pulling on the connector with
approximately 8kg (20lbs) of force. Use the hot air gun to shrink the tubing onto the
connection.
8.
Test the modified MX8031 extension cable. Ensure you have an open circuit between
the center pin and the connector case. Ensure you have continuity between the
cases at both ends of the cable. Ensure you have continuity between the pin at one
end and center conductor at the other end. If any of these checks are not true you
must do the connection over again.
9. Ensure you program the Metrix DPS unit (MX2033 or MX2034) for the closest meter
electrical length that you now have. Verify the system is linear. Perform calibration if
necessary. For instructions on how to modify the Metrix DPS configuration, from the
Factory Purchased configuration go to the below link.
http://www.metrixvibration.com/tenants/metrix/documents/100545.pdf