Chapter II 

28  

Portable digital multimeter 

5.3. DC 

current 

 

Selector 

switch 

Accuracy 

Max. voltage 

drop 

Protection 

Fuses 

 

Resolution 

5 mA 

1 µA 

50 mA 

700 mV 

10 µA 

500 mA 

0.2 % R + 2 D 

1.5 V 

F1 + F2 

100 µA 

10 A

∗∗

 

0.5 % R + 5 D 

500 mV 

600 V

RMS

 

F2

10 

mA

∗ 

refer to fuse specifications section 6.1.1. 

∗∗ 

acceptance overload: 20 A during 30 s. max. observing a break > 5 min between 2 tests  

Number of count :  5000  
Additional error in Pk +/- mode for a pulse width of 

 1 ms : 1 % R 

±

 50 D 

For  measurements performed on alternative signals, the selected range must tally with the max. 
value of the signal peak. 

5.4. 

AC currents (AC and AC + DC) 

 

Accuracy 

40 Hz to  

5 kHz 

5 kHz to 

20 kHz 

20 kHz 

to 30kHz

Ranges 

5 % to 100 % of range 

Protect. 

Fuses

 

Resol. 

Max. 

Crest 

Max. 

voltage 

drop  

5 mA 

1%R+3D 

1.5% typ. 4% typ.

1 µA 

10 mA

50 mA 

1.5% typ. 4% typ.

10 µA 

100 mA

700 mV 

500 mA 

1%R +3 D 

1.5% typ. 4% typ.

F1 + F2

100 µA 

1 A 

1.5 V 

10 A

∗∗

 

1.5%R+3 D 

Æ

 2 kHz 

1.5% typ.

/////////// 

600  

V

RMS

 

F2 

10 mA 

30 A 

500 mV 

 refer to fuse specifications section 6.1.1. 

∗∗ 

acceptance overload: 20 A during 30 s. max. observing a break > 5 min between 2 tests  

Specification : 

from 5 % to 100 % of the range 

Number of count : 

5000 

Additional error according to crest factor :  

0 %  for a crest factor < 1.5 
1 %  for a crest factor of 1.5 to 2 
4 %  for a crest factor of 2 to 3 

Additional error in I

AC+DC

 with a direct current at input : 1 % 

5.5. 

Resistance / Continuity 

  

Ranges

Accuracy

Measur. 

current 

Protection

 

Resolution 

500

 Ω/

0.1% R + 5 D 

1 mA 

100 m

Ω

 

5

 

k

Ω 

100 µA 

Ω

 

50 k

Ω 

10 µA 

10

 Ω

 

500 k

Ω 

0.1% R + 3 D 

1 µA 

100 

Ω

 

5 M

Ω∗∗ 

0.3% R + 3 D 

100 nA 

1k 

Ω

 

50 M

Ω∗∗ 

1% R + 3 D 

50 nA 

600 V

RMS

 

10 k

 Ω

 

 Overload protection can be reset automatically. 

∗∗ 

The use of very short and shielded test leads is highly recommended for measurements in 

this range (> 1 M

Ω

).  

Number of count : 

5000 

Range selection : 

automatic or manual (set in continuity mode) 

Maximum open circuit voltage : 

7 V 

Detection threshold in continuity mode : 

10 

Ω

 to 20

 Ω

 

Response time in continuity mode : 

1 ms 

Содержание MX 58HD

Страница 1: ...istruzioni Manual de instrucciones ENGLISH page 16 Chapter II Office Jl Radin Inten II No 62 Duren Sawit Jakarta 13440 Indonesia Workshop Jl Pahlawan Revolusi No 22B Jakarta 13430 Indonesia Phone 021...

Страница 2: ...Multim tre digital portable 8 9 7 10 11 12 13 5 20 6 19 18 17 16 15 14 1 2 3 4...

Страница 3: ...14 Widerstandsmessung 5 Multimeter Einschalten Zweitfunktionen 15 Kapazit tsmessung 6 Bereichsumschaltung 16 Strommessung bis 5 mA 7 Relativ Messung 17 Strommessung bis 50 mA 8 Spitzenwertmessung 18...

Страница 4: ...test leads 21 3 2 Switching on the instrument 21 3 3 Switching off the instrument 21 3 4 Special configurations 21 3 5 Multimeter maintenance 22 4 FUNCTIONAL DESCRIPTION 23 4 1 SEL ON key 23 4 2 RANGE...

Страница 5: ...rformed on circuits not directly connected to mains E g protected electronic circuits CAT II Measurement category II is for measurements performed on circuits directly connected to the low voltage ins...

Страница 6: ...rent and make sure not to be loaded with static electricity which may destroy internal components Fuses must be replaced with fuses of the same rating and type Any adjustment maintenance or repair wor...

Страница 7: ...related software without the explicit authorization of the manufacturer Repairs necessitated by attempts to repair or maintain the product made by a person not approved by the manufacturer Adaptation...

Страница 8: ...threshold voltage frequencies duty cycles 2 2 Keypad A six key keypad lets you select the autoranging mode RANGE key freeze a value HOLD key measure fast peaks Pk key set the measurement relative to...

Страница 9: ...lly by returning the selector switch to the OFF position or automatically after approximately half an hour if no key is pressed or the switch is not operated Note Automatic shutdown of the instrument...

Страница 10: ...n appears on the display the instrument still has approximately 50 hours of operation but specifications can no longer be guaranteed Replace the battery 3 5 3 Replacing the battery or fuses Disconnect...

Страница 11: ...sed to access secondary functions associated with the selector switch positions The flowcharts below define these various functions 4 1 1 VAC position 4 1 2 mV position SEL ON SEL ON DC voltage measur...

Страница 12: ...imeter 4 1 3 VDC position 4 1 4 position 4 1 5 position SEL ON SEL ON AC DC voltage measurement DC voltage measurement SEL ON SEL ON Continuity test Resistance measurement SEL ON SEL ON Diode voltage...

Страница 13: ...RANGE key is used to change from one range voltage to the next range The new range is then displayed for 2 seconds 4 3 REL key Short press Mode REL the last value measured becomes the reference value...

Страница 14: ...urement is stable When you remove the probes a second audible signal indicates that this stable value displayed has been stored 4 6 PRINT key This key operates with the optional serial interface for p...

Страница 15: ...Serial mode rejection at 50 and 60 Hz better than 60 dB Additional error in Pk mode for a pulse of 1 ms 1 R 50 D For measurements performed on alternative signals the selected range must tally with t...

Страница 16: ...1 F2 100 A 1 A 1 5 V 10 A 1 5 R 3 D 2 kHz 1 5 typ 600 VRMS F2 10 mA 30 A 500 mV refer to fuse specifications section 6 1 1 acceptance overload 20 A during 30 s max observing a break 5 min between 2 te...

Страница 17: ...threshold voltage measurement Measurable voltages 0 to 2 V Measurement current 1 mA typical Resolution 1 mV Protection 600 VRMS can be reset automatically 5 8 Frequencies Selector switch setting VAC S...

Страница 18: ...0 M 0 to 70 from 40 C to 50 C 60 above 50 C Casing and circuit self extinguishing materials Operating quality IEC 359 Electromagnetic compatibility Emission Immunity EN 61326 1 Max influence in electr...

Страница 19: ...ps 0 5 to 240 AAC 20 mm MN09 0 1 to 1200 AAC 52 mm CI03 0 5 to 1000 AAC 39 mm PAC 20 0 5 to 1000 AAC 39 mm PAC 21 Shunts 30 A 300 mV HA030 1 50 A 100 mV HA050 Miscellaneous Set of RS232 serial link SX...

Отзывы: