Chapter II
Benchtop multimeter
25
5.2.
AC voltages (AC and AC+DC)
Selector
Ranges
Accuracy
Input
Protection
Reso-
switch
position
DC*
40 Hz
to
1 kHz
1 kHz
to
4 kHz
4 kHz
to
10 kHz
10 kHz
to
20 kHz
20 kHz
to
30 kHz
impedance
lution
5 % to 100 % full scale
10 % to 100 % full scale
mV
500 mV *
4%R+3 D
////////////
////////////
/////////////
10 M
Ω
±
1100
V
PK
*
100 µV
V
AC
5 V
11 M
Ω
±
1100
V
PK
1 mV
or
50 V
1 % R + 3 D
2 % R + 3 D
3 % R + 3 D
10 M
Ω
±
1100
V
PK
10 mV
V
DC
500 V
10 M
Ω
±
1100
V
PK
100 mV
+ SEL
750 V
////////////
////////////
/////////////
/////////////
10 M
Ω
±
1100
V
PK
1 V
* AC+DC only
** 1 mn max
Number of points:
5 000
Range selection:
Automatic or manual for the 5 V, 50 V, 500 V,
750 V ranges
Common mode rejection:
At 50 and 60 Hz, better than 80 dB
Additional error according to peak factor :
0 % for a peak factor < 1.5
1 % for a peak factor of 1.5 to 2
4 % for a peak factor of 2 to 3
5.3. DC
current
Selector switch
position
Accuracy
Max voltage drop
Protection
Fuses *
Resolution
5 mA
0.2 %R + 2 D
700 mV
600 V
RMS
F1 + F2
100 nA
50 mA
0.2 %R + 2 D
700 mV
600 V
RMS
F1 + F2
1 µA
500 mA
0.2 %R + 2 D
1.5 V
600 V
RMS
F1 + F2
10 µA
10 A
0.5 %R + 5 D
500 mV
600 V
RMS
F2
1 mA
* Refer to fuse specifications section 6.1.1
Number of points:
50 000 (or 5 000 - refer to § 3.3.)
Additional error in Pk +/- mode for a pulse width of
≥
1 ms : 1% R
±
50 D
For measurements performed on alternative signals, the selected range must tally with the
max. value of the signal peak.
5.4.
AC currents (AC and AC+DC)
Selectorswitch
position
Accuracy
Max
voltage
drop
Protection
Fuses *
Resolution
Max.
peak
40 Hz to 5 kHz
5 mA
700 mV
600 VRMS
F1 + F2
1 µA
10 mA
50 mA
1 % R + 3 D
700 mV
600 VRMS
F1 + F2
10 µA
10 0mA
500 mA
1.5 V
600 VRMS
F1 + F2
100 µA
1 A
10 A
1.5 % R + 3 D
→
2 kHz
500 mV
600 VRMS
F2
10 mA
* Refer to fuse specifications section 6.1.1
Specification :
from 5% to 100% of the range
Number of points:
5 000
Additional error according to peak factor: 0% for a peak factor < 1.5
1% for a peak factor of 1.5 to 2
4% for a peak factor of 2 to 3
Additional error in I
AC+DC
with a direct current at input : 1 %
Содержание MX 553
Страница 2: ...Multim tre de table...
Страница 65: ...Cap tulo IV Pol metro de sobremesa 61...