How are the various quantities measured? 

20 

Benchtop multimeters, 6000 and 60000 points 

 

How are the various quantities measured? (continued) 

6. Diode test 

1.  Set the switch to 

 . 

2. Press 

 twice; the “

” symbol is displayed. 

3.  Connect the black lead to the “COM” terminal and the red lead to “

+

”. 

4.  Place the test probes on the terminals of the component: 

 

5.  Read the measured threshold voltage of the junction indicated  

on the display unit. 

6. “

O.L

” is displayed, if the circuit is open or the threshold of the diode >3V.  

 

7. Capacitance  
measurement 

1.  Set the switch to 

 . 

2.  Connect the black lead to the “COM” terminal and the red lead to “

+

”. 

3.  Place the test probes on the terminals of the component: 

 

4.   Read the measurement value indicated on the display unit. 

 

O.L

” is displayed, if the value to be measured exceeds the capacitance of 

the range.  
 

O.L

” is displayed, if the capacitor is short-circuited. 

 

¾

  For high values, the measurement cycle includes the display of "run" with 

a "chaser" decimal point. This means that acquisition is in progress; wait 
for the display of the digital result. 

 

 

"Run" is displayed immediately, if the previous measurement 

 

was in a small range. 

 

¾

  The prior discharge of very high capacitances helps shorten the 

measurement time. 

 

Содержание MX 5006

Страница 1: ...60 0 00 0 00 0 0p p pt t ts s s M M MX X X 5 5 50 0 06 6 60 0 0 6 6 60 0 00 0 00 0 00 0 0p p pt t ts s s U U Us s se e er r r s s s M M Ma a an n nu u ua a al l l P le Test et Mesure CHAUVIN ARNOUX Pa...

Страница 2: ...back 7 Functional description 8 Display unit 8 Switch 10 Keypad 11 Preparation for use 16 How are the various quantities measured 17 1 Voltage measurement 17 2 Current measurement 18 3 Frequency measu...

Страница 3: ...the instrument is the responsibility of the system integrator It can be used for measurements on 1000V CAT III and 600V CAT IV circuits and can be supplied by a 230V 10 AC network 230V CAT II However...

Страница 4: ...g a measuring instrument a lead or an accessory belonging to a lower measurement or voltage category derates the resulting system instrument leads accessories to the lowest measurement category and or...

Страница 5: ...from line power and from the measurement circuits and make sure that you are not charged with static electricity which might destroy internal components An adjustment maintenance or repair of the live...

Страница 6: ...t leave the COM socket open 3 The display unit must indicate a result 2 if not replace the fuse Access to and location of the fuse by opening the housing Current protection fuse 11A 1000V 20kA 10x38 P...

Страница 7: ...of the instruments Front panel MX 5006 LCD 6000pts Keypad Switch Terminal block MX 5060 LCD 60000pts Keypad Switch Terminal block Rear panel MX 5006 Reminders Switch Power connector ON OFF 110V or 230...

Страница 8: ...VAC AC voltage measurement VAC DC DC or AC DC voltage measurement at high impedance V A Current measurement A Hz Frequency measurement Resistance measurement F Capacitance measurement T Temperature me...

Страница 9: ...peak value run r un ru n Capacitance meter acquisition in progress Frequency measurement impossible O L Overshoot of the measurement capacities V Volt Hz Hertz F Farad C F Degree Celsius degree Fahren...

Страница 10: ...one measurement function to another deactivates the key if the HOLD mode was selected The switch has 10 positions 3 2 1 4 5 6 7 8 9 10 1 OFF mode Switches the multimeter off 2 AC voltage measurement...

Страница 11: ...of the measurement range The range defines the maximum measurement span the instrument can cover The Auto Range mode is activated as default Display of the MAX MIN PEAK or PEAK mode MAX and MIN displa...

Страница 12: ...lay Acquisition continues as a background task In the MAX MIN PEAK mode when HOLD is active the blinking of the MAX MIN PEAK symbol indicates that acquisition continues as a background task On off swi...

Страница 13: ...the key The measured signal changes to 250V 50Hz for the MIN value Second press on the key for the PEAK value Third press on the key for the PEAK value Fourth press on the key Present value of the si...

Страница 14: ...5V REL 1 5V 1V 0 5V De activation of the REL mode by a press on the key A long press on the key erases the reference value and exits from the REL mode Short press in the REL mode on the key Present va...

Страница 15: ...inations according to the type of measurement Range Type of measurement Max Min Peak REL Auto Manu HOLD Voltage VLowZ Voltage VAC Voltage VAC DC Current AAC AAC DC 9 9 9 in REL only 9 9 9 9 Voltage VD...

Страница 16: ...carry currents to earth Powering up down Use the power On Off switch on the back of the instrument to power it up A check light on the front of the instrument indicates that it is powered up Switchin...

Страница 17: ...ement exceeds 600V 1 Set the switch to or or 2 Select AC DC or DC coupling of the signal by pressing the default coupling is AC DC Depending on what you select the screen displays DC or AC DC 3 Connec...

Страница 18: ...the red lead to A 4 Place the test probes in series between the source and the load 5 Read the measurement value indicated on the display unit O L is displayed if I 20A 6 As default the 2nd display u...

Страница 19: ...nd the red lead to 3 Place the test probes on the terminals of the component Remark Resistance measurements must be made with power off However while the presence of a voltage will prevent or throw of...

Страница 20: ...citance measurement 1 Set the switch to 2 Connect the black lead to the COM terminal and the red lead to 3 Place the test probes on the terminals of the component 4 Read the measurement value indicate...

Страница 21: ...ult on the main display unit is C 3 Connect the temperature probe K thermocouple to the COM and terminals bearing the polarity in mind 4 Read the measurement value indicated on the display unit If O L...

Страница 22: ...the COM terminal and the red lead to 4 Place the test probes between two phases of the circuit to be measured 5 Read the measurement values indicated on the display unit voltage and frequency O L is d...

Страница 23: ...bove 1050V and OL below 1050V AC voltage VLowZ AC RMS The bandwidth is reduced to 300Hz if the filter is activated The frequency measurement is made like a measurement in a 300Hz PB 1 See the typical...

Страница 24: ...0V or above 1050VRMS 3 From 1kHz the measurement must exceed 15 of the range Secondary measurements and displays FREQ AC coupling MAX MIN PEAK AC and DC voltage AC DC TRMS Range Operatin g range Speci...

Страница 25: ...to 600 0mA 0 2 to 600 0mA 0 1mA 0 8 L 2 D 0 58mV mA 6A 0 to 6 000A 0 200 to 6 000A 0 001A 0 8 L 3 D 0 05V A 10A 20A 0 to 20 00A 0 20 to 20 00A 0 01A 0 8 L 2 D 0 05V A 11A 1000V fuse 20kA AC current A...

Страница 26: ...ents and displays FREQ AC coupling MAX MIN PEAK Acceptable overload 10A to 20A for 30s max with a pause of 5min between 2 measurements Ambient temp 35 C max Additional uncertainty with the 300Hz filte...

Страница 27: ...ime 100ms Range Resolution Uncertainty Open circuit voltage Measurement current 600 0 1 Audible signal triggered 30 5 5V 1 1mA Diode Test Protection 1414Vpk Range Resolution Uncertainty Open circuit v...

Страница 28: ...nty 200 0 C to 200 0 C 60 0 C to 200 0 C 0 1 C 0 5 L 2 C low 328 0 F to 392 0 F 76 0 F to 392 0 F 0 1 F 0 5 L 4 F 200 C to 1200 C 60 C to 1200 C 1 C 0 5 L 2 C high 328 F to 2192 F 76 F to 2192 F 1 F 0...

Страница 29: ...L 0 1 D 1 C 0 05 L 0 1 D 1 C VAC VAC DC VLowz 0 15 L 0 1 D 1 C ADC 0 05 L 0 1 D 1 C 0 1 L 0 1 D 1 C AAC and AAC DC 0 08 L 0 1 D 1 C 0 12 L 0 1 D 1 C 0 01 L 0 1 D 1 C 0 1 L 1 C 0 05 L 1 C 0 1 L 1 C 60M...

Страница 30: ...0 07 L 25 D 1 This range is accessible only with the key Input impedance approx 10 6M 50pF 2 The display indicates OL above 1050V and OL below 1050V AC voltage VLowZ AC RMS The PB is reduced to 300Hz...

Страница 31: ...low 1050V or above 1050VRMS 4 From 1kHz the measurement must exceed 15 of the range Secondary measurements and displays FREQ AC coupling MAX MIN PEAK 60mV range Measuring a strong current or measuring...

Страница 32: ...20 D 0 58mV mA 6A 0 to 6 0000 A 0 2000 to 6 0000A 0 0001A 0 8 L 20 D 0 05 V A 10A 20A 0 to 20 000A 0 200 to 20 000A 0 001A 0 8 L 20 D 0 05 V A Fuse 11A 1000V 20kA AC current AC RMS Range Operating ra...

Страница 33: ...0 0 A 60 to 6000 0 A 0 1 A 1 L 0 08 x F kHz 1 L 25 D 2 6 to 5mA 25mV mA 60mA 0 to 60 00mA 6 000 to 60 000mA 0 001mA 2 6 to 50mA 3mV mA 600mA 0 to 600 00mA 60 00 to 600 00mA 0 01mA 1 L 0 08 x F kHz 1 L...

Страница 34: ...dary display Particular reference conditions 150mV U 600V 0 15A I 10A Max frequency measurable in volts 60kHz Max frequency measurable in amperes 60kHz When the switch is set to VLowZ Volts or Ampere...

Страница 35: ...0 F 0 001 F 1 L 5 D 12 6 A 0 125 s F 60 F 0 to 60 00 F 0 to 60 00 F 0 01 F 1 L 5 D 126 6 A 0 125 s F 600 F 0 to 600 0 F 0 to 600 0 F 0 1 F 3 L 5 D 1mA 0 125 s F 6mF 0 to 6 000mF 0 to 6 000mF 1 F 4 L 5...

Страница 36: ...upper limit on the temperature display other than the 6000 D of the display unit PEAK PEAK Add 1 L 30 D to obtain the accuracy corresponding to the function and the range Fmax 1kHz 1ms MAX MIN Add 0 2...

Страница 37: ...05 L 0 1 D 1 C VAC VAC DC VLowZ 0 15 L 0 1 D 1 C ADC 0 05 L 0 1 D 1 C 0 1 L 0 1 D 1 C AAC and AAC DC 0 08 L 0 1 D 1 C 0 12 L 0 1 D 1 C 0 01 L 0 1 D 1 C 0 1 L 1 C 0 05 L 1 C 0 1 L 1 C 60 M 0 3 L 1 C n...

Страница 38: ...Display The refresh rate of the display unit is 200ms Safety According to NF EN 61010 1 Insulation class 2 Degree of pollution 2 Use indoor Altitude 2000m Measurement category of the measurements inp...

Страница 39: ...es on CD ROM Getting started guide EU line power cord Lead 1 5m straight straight red Lead 1 5m straight straight black Test probe CAT IV 1kV red Test probe CAT IV 1kV black MX 5060 USB cord optional...

Отзывы: