Chapter II
5000-count portable digital multimeter, with current input
25
6. TECHNICAL
SPECIFICATIONS
6.1. General
Only the values assigned tolerances or given limits are guaranteed values. Values without
tolerances are given for information only, without guarantee (standard NFC 42670). The
measurement errors must be considered in an environment of reference temperature (refer to
§ 6.2.11).
It is essential that all measuring instruments are regularly calibrated.
6.2. Characteristics
Accuracy is
±
[% reading (R) + number of digits (D)].
{Accuracy : ‘’n %R + n D’’ means ‘’n % of the r n digits’’ }
6.2.1. DC voltages
Selector switch
position
Ranges
Accuracy
Input impedance
Protection
(∗)
Resolution
ADP
DC
500 mV
DC
10 M
Ω
0.1 mV
DC
5 V
DC
11 M
Ω
±
1100 V
PEAK
1 mV
DC
V
DC
50 V
DC
0,3 % R + 2 D
775 Vrms
10 mV
DC
500 V
DC
10 M
Ω
100 mV
DC
1000 V
DC
1 V
DC
(∗)
Acceptance permanent max. voltage
Number of points :
5 000
Range selection :
automatic or manual for the 5 V, 50 V,
500 V, 1000 V ranges
Common mode rejection :
at 50 and 60 Hz, better than 120 dB
Serial mode rejection :
at 50 and 60 Hz, better than 60 dB
Intermittent signal sounds and « OL » on the display in case of range overpassing.
For measurements performed on alternative signals, the selected range must tally with the
max. value of the signal peak.
Additional error in Pk +/- mode for a pulse width of
≥
1 ms : 1 % R
±
50 D
6.2.2. AC voltages (AC and AC+DC)
Switch
Range
Input
Protection
Accuracy
Resolution
position
impedance
(*)
40Hz to 1 kHz
1 to 4 kHz
4 to 20 kHz 20 to 50 kHz 50 to 100 kHz
5 V
AC
1 mV
AC
VLOW Z
50 V
AC
500 k
Ω
± 1100 Vpk
1 % L + 3 D
1 % L + 3 D
2 % L + 3 D 3 % L + 3 D
5 % L + 3 D
10 mV
AC
500 V
AC
775 Vrms
100 mV
AC
750 V
AC
-
-
-
-
1 mV
AC
ADPAC+DC
500 mV
AC+DC
10 M
Ω
//100 pF
± 1100 Vpk
775 Vrms
1.5 % L + 3 D
-
-
-
-
0.1 mV
AC+DC
5 V
AC+DC
11 M
Ω
1 mV
AC+DC
VAC+DC
50 V
AC+DC
± 1100 Vpk
1 % L + 3 D
1 % L + 3 D
2 % L + 3 D 3 % L + 3 D
5 % L + 3 D
10 mV
AC+DC
500 V
AC+DC
10 M
Ω
775 Vrms
100 mV
AC+DC
750 V
AC+DC
-
-
-
-
1 V
AC+DC
(∗)
Acceptance permanent max. voltage
Specifications applicable from 10 % to 100 % of the range for a sinusoidal signal in band 40 Hz
to 20 kHz and from 20 % to 100 % of the range in band 20 kHz to 100 kHz.
Number of points :
5 000
Range selection :
automatic or manual for the 5 V, 50 V,
500 V, 750 V ranges
Common mode rejection :
at 50 and 60 Hz, better than 60 dB
Additional error according to crest factor :
0.5 % for a crest factor of 2 to 3
1 % for a crest factor of 3 to 6
Intermittent signal sounds and « OL » on the display in case of range overpassing.