background image

Chapter II 

5000-count portable digital multimeter, with current input 

25 

5.2.6.  Capacitance 

F

 

Note 

Discharge all capacitors before taking measurements. 

 

Selector switch 

position 

Ranges 

Accuracy 

Measurement 

current 

Max measurement 

time 

Protection 

(∗)

 

Resolution 

 

50 nF 

 

100 nA 

< 1 s 

 

10 pF 

 

500 nF 

 

1 µA 

< 1 s 

 

100 pF 

 

5 µF 

 

10 µA 

< 1 s 

 

1 nF 

 

50 µF 

1 %R + 2 D 

100 µA 

< 1 s 

600 V

RMS

 

10 nF 

 

500 µF 

 

1 mA 

< 2 s 

 

100 nF 

 

5000 µF 

 

1 mA 

 3 s/mF 

 

1 µF 

 

50 mF 

 

1 mA 

 3 s/mF 

 

10 µF 

(∗)

 Overload protection can be reset automatically. 

 

To reduce the influence of the ambient electromagnetic fields on the measurement, it is 
recommended to use shielded cables for measurements in 50 µF range. 
Number of points : 

5 000 

Range selection : 

automatic or manual 

Maximum open circuit voltage : 

4 V 

 

5.2.7.  Diode threshold voltage measurement 

 

Measurable voltages : 

0 to 1.999 V 

Measurement current : 

1 mA ± 20 %  

Resolution : 

1 mV 

Protection : 

600 V

RMS

, can be reset automatically 

 

5.2.8.  Frequencies 

 

Selector switch 

position 

Ranges 

Accuracy 

Input 

 impedance 

Protection 

Resolution 

 

0.62 to 5 Hz 

 

 

 

0.0001 Hz 

 

5 to 50 Hz 

 

 

 

0.001 Hz 

FREQ 

50 to 500 Hz 

0.03 %R + 1 D 

10 M

   

600 V

RMS

 

0.01 Hz 

 

500 to 5 kHz 

 

 

 

0.1 Hz 

 

5 to 50 kHz 

 

 

 

1 Hz 

 

50 to 500 kHz 

0.05 %R + 1 D 

 

 

10 Hz 

 

Number of points : 

50 000 

Range selection : 

automatic 

 

Frequency band 

Sinus 

Square 

10 Hz to 100 kHz 

100 mVeff. 

200 mVpp 

100 kHz to 500 kHz 

1 Veff. 

1 Vpp 

 

 

5.2.9.  Safety 

 

IEC 61010-1 + A1 + A2, 1995 
Insulation : class 2 
Pollution degree : 2 
Indoor use, altitude < 2000 m 
Installation category of input :   

CAT 

III,

 600 V max. to earth 

 

Sensitivity : 
:::

:

 

Содержание MX 24B

Страница 1: ...IMETER DIGITAL MULTIMETER MULTIMETRO DIGITALE MULTIMETRO DIGITAL Notice de fonctionnement page 1 Chapitre User s manual page 14 Chapter Bedienungsanleitung Seite 28 Kapitel Libretto d istruzioni pagin...

Страница 2: ...Multim tre digital portable 5000 points avec entr e courant 5 6 18 17 16 15 14 13 12 11 7 1 2 3 4 19 20 21 22 8 9 10 500mA MX 24B 500mADC...

Страница 3: ...nfiguration 19 3 5 Multimeter maintenance 19 3 5 1 Fuse self test 19 3 5 2 Battery self test 20 3 5 3 Replacing the battery or fuses 20 3 5 4 Cleaning 20 3 5 5 Storage 20 4 FUNCTIONAL DESCRIPTION 21 4...

Страница 4: ...CAT III The circuits of CAT III are power supply circuits of power equipment with important transient overvoltages Example fixed installation or industrial equipment CAT IV The circuits of CAT IV may...

Страница 5: ...replaced with fuses of the same rating and type Any adjustment maintenance or repair work carried out on the multimeter while it is live should be carried out only by appropriately qualified personnel...

Страница 6: ...misuse of the equipment or use in association with incompatible equipment 2 Modification of the equipment or any related software without the explicit authorization of the manufacturer 3 Repairs nece...

Страница 7: ...ect the autoranging mode RANGE AC DC key store a value MEM key activate the MIN MAX AVG detection mode MIN MAX key select a function derived from the main function or switch on the multimeter again af...

Страница 8: ...operated F Note For user safety automatic shutdown is also disabled when a measured magnitude voltage current present at the input exceeds dangerous level indicator displayed 3 4 Special configuratio...

Страница 9: ...Use the following procedure 1 Disconnect test leads from any inputs terminals 2 Using a appropriate tool make slide the case bottom of the instrument 3 Replace the battery or fuse respect the value a...

Страница 10: ...multimeter again after an automatic shutdown It can also be used to access secondary functions associated with the selector switch positions mode AC DC The flowcharts below define these various funct...

Страница 11: ...ond short press returns the multimeter to normal mode The MEM sign appears on the display The MEM mode is avaible on all measurement Long press AUTO MEM Accesses or quits the autostore mode The MEM si...

Страница 12: ...1000 V ranges Common mode rejection 500 mV range at 50 and 60 Hz better than 120 dB Serial mode rejection at 50 and 60 Hz better than 60 dB Intermittent signal sounds and OL on the display in case of...

Страница 13: ...5 3 Acceptance 20 A overload during 30 s max observing a 5 min break at least between two measurements Specifications applicable from 5 to 100 of the range for a sinusoidal signal in band 40 Hz to 1 k...

Страница 14: ...of points 5 000 Range selection automatic or manual Maximum open circuit voltage 4 V 5 2 7 Diode threshold voltage measurement Measurable voltages 0 to 1 999 V Measurement current 1 mA 20 Resolution...

Страница 15: ...measurements s Bargraph 20 measurements s 5 2 11 Environment Reference temperature 23 C 5 C Limit range of operation 5 C to 45 C Rated range of use 10 C to 55 C Storage temperature range 20 C to 70 C...

Страница 16: ...K thermocouple 1 mV C general purpose and surface type 25 C to 350 C HK0210N Optical tachometer 100 rpm to 60 000 rpm HA1237 Current clamps Range 200 AAC 1 AAC 1 mVAC max cable 20 mm connection socket...

Страница 17: ...g des Multimeters 33 3 5 1 Selbsttest der Sicherungen 33 3 5 2 Selbsttest der Batterie 34 3 5 3 Ersetzen der Batterie und der Sicherungen 34 3 5 4 Pflege 34 3 5 5 Lagerung 34 4 FUNKTIONSBESCHREIBUNG 3...

Отзывы: