IX 3131B
II - 45
Test frequency : 10 kHz
Accuracy (DF < 0.5)
Range
Max.
display
Inductance
Dissipation Factor
Notes
1000 mH
999.9 mH
2.0 % +
(Lx/10000) % + 8
*N4
1.5%+100/Lx+10
-
200 mH
199.99 mH
0.6 % +
(Lx/10000) % + 8
1.5%+100/Lx+10
-
20 mH
19.999 mH
0.6 % +
(Lx/10000)%+10
2.0%+100/Lx+15
-
2000
µ
H
1999.9
µ
H
1.0 % +
(Lx /10000)% +10
5.0%+100/Lx+20
after short cal.
Notes
1 -
Q Value is the reciprocal of DF.
2 -
This specification is based on the measurement performed
at the test leads.
4 -
Lx = counts of displayed L value. e.g. L = 88.88H, then
Lx = 8888.
5 -
The accuracy is specified for series mode in this range.