Kit Demo for METRIX Oscilloscopes
21
Demo:
with:
DOX2025
DOX2040
DOX2100
DOX2xxx
Test Signal
no. 10: Digital frame+Fault
Nature
Digital frame with a recurrent fault
Specs
F, square wave
5 MHz, Vpp
1.8V-L+ fault
7ns
Oscilloscope Settings
25 or 50ns/div., then 250ns/div-MAIN = 500mV/div., DC coupling
Trigger
DC coupling on MAIN, level
250mV
Modes
Select "Repetitive Signal" (Horiz menu)
Objectives
Using triggering on pulse width
Using the LongMem and Delayed mode
a) Display the signal (possible using the "Autoset" key), then set the parameters as indicated below.
It can be seen that the display is not stable.
Set the triggering to pulse width < 20ns and adjust the triggering level to close to the low level of the pulse in
order to trigger on the fault.