![Metrix DOX 2 Series Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/metrix/dox-2-series/dox-2-series_user-manual_1779215012.webp)
10
Kit Demo for METRIX Oscilloscopes
Demo:
with:
DOX2025
DOX2040
DOX2100
DOX2xxx
Test Signal
no. 4: Data+CS train
Nature
2 signals representing a digital frame (data) and a CS (chip select)
Specs
Vpp
3.4V-F
40kHz (data)-F
1.5kHz (CS)
Oscilloscope Settings
100µs/div.-MAIN = 1 V/div.-AUX = 1 V/div.
Trigger
on BNC AUX = CH2
Modes
Triggered mode preferable
Objectives
Triggering on pulses
a) Adjust the oscilloscope to display simply the 2 signals (time base, sensitivities and
triggering source on the
BNC AUX = CH2).
Attention, with this type of signal, "Autoset" operation may be aleatory.
b) We are now going to demonstrate the utility of the pulse width triggers.
The example chosen will serve to synchronize to the chip select signal of the data frame.
We are going to trigger by turns on the width of the high level, then of the low level, of the "positive" pulse. In
the first case, triggering will be on the negative-going edge of the chip select and in the second case it will be
on the positive-going edge.
Содержание DOX 2 Series
Страница 1: ...X04128A02 Ed 1 09 15 Demonstration SET for METRIX DOX 2xxx Oscilloscopes User s guide...
Страница 9: ...Kit Demo for METRIX Oscilloscopes 7 FFT of the signal on channel CH2...
Страница 16: ...14 Kit Demo for METRIX Oscilloscopes Zoom by 1000 in LongMem mode...
Страница 27: ...Kit Demo for METRIX Oscilloscopes 25...
Страница 45: ...Kit Demo for METRIX Oscilloscopes 43...
Страница 48: ...46 Kit Demo for METRIX Oscilloscopes Remark the 70 kHz and 90 kHz components have the same amplitude...