Between the light-entering prism (1) and the refracting prism (2) there is a small gap, in which the liquid to be 
measured is placed. When a light beam (natural light or incandescent light) enters the light-entering prism (1), a 
diffuse reflection will be caused on its frosted surface; therefore, incident rays with various angles in the liquid to 
be measured will, via the reflecting prism (2), form a light beam with reflection angles larger than the angle of 
emergency.  
The swinging reflecting mirror (3) guides the light beam into the Amici prisms unit (4), which consists of a pair of 
Amici prisms, and plays a function of using a changeable dispersion to offset against the dispersion produced by 
the refracting prism to different substance to be measured. Then the telescope objectives unit (5) images the 
separatrix on the dividing scale (7), on which the graticule can be seen through the eyepiece (8), an image as 
shown on the upper part of Figure 5. 
lluminated by light beam via the condensing lens (12), the scale plate (11) links with the swinging 
reflecting mirror (3) to form an integral whole, and gyrates around the graduation centre.  
Through the reflecting mirror (10) and the reading lens (9), the parallel prism (6) images the indicating 
values of the refractive indexes from different positions of the scale plate onto the dividing scale (7) 
(referred to the image on the lower part of Figure 5) 
 

 

 

4.2 Assembly (see figures 6 and 7) 

 
The base (14) is a cradle for the instrument, on which the casing (17) is mounted. All optical components and the 
main structure are enclosed in the casing except the eyepiece and prisms. The prism unit is mounted on the 
casing, consisting of light-entering prism, refracting prism and prism base. Two prisms are fixed in the prism 
holder with a special binding agent. (5) The light-entering prism holder and (11) the refracting prism holder, 
linked by the rotation shaft (2) the light-entering prism can be opened or closed. When the two prism holders 
are tightly closed and locked by the hand wheel (10), a uniform gap is left between the two surfaces of the 
above prisms and the liquid to be measured should be filled in the gap. (3) the hood, (18) is the adapter 
connector for four thermostats, (4) is the thermometer, and (13) is the thermometer holder, which can be 
connected to the thermometer with an emulsive tube. (1) is the reflecting mirror, (8) is the eyepiece, (9) is the 
cover plate, (15) is the refractive index graduation adjusting hand wheel, (6) is the dispersion adjusting hand 
wheel, (7) is the dispersion graduation ring, and (12) is the condensing lens for illuminating the dial scale. 

 

 

5.

 

Operation 

 

5.1 Preparation 

 

5.1.1

 Before performing the measurement, use distilled water or standard sample to check the readings. In case 

of using standard sample, put one or two drops of naphthalene bromide onto the polished surface of the 
refracting prism, then make it contact with the polished surface of the standard sample. When the indication of 
the field of view is higher than the value of the standard sample, observe whether the separatrix is in the centre 
of the graticule. If there is a deviation, slightly turn the screw inside the small hole (16) of Figure 7 with a screw 
driver, making the objective to offset until the separatrix moves to the centre of the graticule. Repeat the 
observation and correction until to minimize the indication of the initial error (including the collimating error of 
the operator). After the completion of the correction, this location should be kept unchanged throughout the 
process of measurement. 
Generally, the instrument needs not to be checked in the routine measurement. In case of having doubted of the 
indication of the measured refractive index, perform the checking procedure as specified in the above method 
to make clear whether there is an initial error and make a correction if necessary. 

Содержание RST111

Страница 1: ...ABBE Analogic refractometer RST111 Please read the User Manual carefully before use and follow all operating and safety instructions user manual english...

Страница 2: ...are representative with the following information Serial number Description of problem Your contact information Warranty This instrument is warranted to be free from defects in materials and workmansh...

Страница 3: ...perature between 0 and 70 Both the refractive index and the average dispersion are the important optical constants through which we can understand the optical performance purity concentration as well...

Страница 4: ...0 Hand wheel 16 Adjusting cavity 5 Light entering prism holder 11 Refraction prism holder 17 Casing 6 Dispersion adjusting wheel 12 Condensing lens 18 Adapter connector for 4 thermostats 1 Light enter...

Страница 5: ...and 11 the refracting prism holder linked by the rotation shaft 2 the light entering prism can be opened or closed When the two prism holders are tightly closed and locked by the hand wheel 10 a unif...

Страница 6: ...x in the field of view of the eyepiece the operating methods of sighting and reading are the same as above 5 2 3 Measurement of translucent solid Stick the polished surface of the translucent solid to...

Страница 7: ...in the sample to be measured When measuring the solid sample prevent the working surface of the refracting prisms from being roughed and scratched 6 4 Regularly keep the instrument clean Strictly pro...

Страница 8: ...8...

Страница 9: ...ake measuring the average dispersion of distilled water as an example When temperature is 20 nD 1 3330 The reading of dispersion value indicated on the scale ring will be Turn in a certain direction T...

Страница 10: ...tomatically Turn off and put the key for four seconds turn on put the key back on When the temperature is higher than 70 C display will show H when is lower than 50 C display will show L C When the ba...

Страница 11: ...desecharse en vertederos De conformidad con la Directiva 2002 96 CE los usuarios de la Uni n Europea de aparatos el ctricos y electr nicos tienen la oportunidad de retornar el instrumento para su eli...

Страница 12: ...www labbox com...

Отзывы: