MI 3290 Earth Analyser
Auto Sequences
®
120
Auto Sequence
® view menu (measurement is selected)
Name of Auto Sequence
Parameters and limits of
selected single test
Header
Single
Test
Options
Figure 12.3: Auto Sequence
® view menu – measurement selected
Options
Selects single test.
on
Opens menu for changing parameters and limits of selected
measurements.
Refer to chapter
10.1.2 Setting parameters and limits of
for more information how to change
measurement parameters and limits.
Indication of Loops
The attached ‘x2’ at the end of single test name indicates that a loop of single tests is programmed.
This means that the marked single test will be carried out as many times as the number behind the
‘x’ indicates. It is possible to exit the loop before, at the end of each individual measurement.