MD 9020 Electrical & Electronic DMM
Specification
17
Low Battery
:
Below approx. 2.4V
E.M.C.
:
Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and EN61000-4-3(1996)
In an RF field of 3V/m:
CAPACITANCE FUNCTION IS NOT SPECIFIED
AC 4.000V range:
Total Accuracy = Specified Ac 700 digits
AC 400.0
A range:
Total Accuracy = Specified Ac 300 digits
Other function ranges: Total Accuracy = Specified Ac 40 digits
Performance above 3V/m is not specified
Sleep Mode Timing
:
Idle for 30 minutes
Sleep Mode Consumption
:
300
A typical for MD 9020;
Dimension:
L186mm X W87mm X H35.5mm; L198mm X W97mm X H55mm with
holster
Weight:
296 gm; 396 gm with holster
Special Features
:
25ms Max Hold; Data Hold; Relative zero mode; Beep-jack™ input
warning; Back-lighted display
Accessories
:
Test leads (pair), batteries installed, user's manual
Optional Accessories
:
AMD 9023 - Banana plug type-K bead probe,
AMD 9024 - Banana pins to type-K socket plug adapter.
ELECTRICAL SPECIFICATIONS
Accuracy is ±(% reading number of digits) or otherwise specified, at 23
O
C ±5
O
C & less than 75%
R.H.
Voltage
RANGE Accuracy
400.0 mV
0.3% + 4d
4.000V, 40.00V, 400.0V 0.5% + 3d
1000V
1.0% + 4d
NMRR: >50dB @ 50/60Hz
CMRR: >120dB @ DC, 50/60Hz, Rs=1k
Input Impedance: 10M
, 30pF nominal
(1000M
for 400.0mV range)
Max Hold (Voltage & Current)
Specified accuracy ± 50 digits for changes > 25ms in duration
AC Voltage
RANGE Accuracy
1)
50Hz -- 500Hz
400.0mV*
4.0% + 5d
4.000V, 40.00V, 400.0V 1.5% + 5d
1000V
4.0% + 5d
CMRR
:
>60dB @ DC to 60Hz, Rs=1k
Input Impedance
: 10M
, 30pF nominal
(
1000M
for 400.0mV range)